A-96.250.671 / 070222
AMI Silica
Default Values
97
11. Default Values
Operation:
Sensors:
Filter Time Const.:........................................................................ 10 s
Hold after Cal.:........................................................................... 300 s
Alarm Relay
.........................................................................same as in Installation
Relay 1 and 2
.........................................................................same as in Installation
Input
.........................................................................same as in Installation
Logger
Logger Interval:.............................................................. Event Driven
Clear Logger:.................................................................................. no
Installation:
Sensor
Ref. Verification:......................................................................... 0.235
Standard: ............................................................................... 100 ppb
Blank....................................................................................... 0.0 ppb
Meas. interval .......................................................................... 10 min
Channels........................................................................................... 1
Channel Selection................................................................... internal
Signal Output 1
Parameter: .................................................................................... Si 1
Current loop:....................................................................... 4 –20 mA
Function:.................................................................................... linear
Scaling: Range low:................................................................ 0.0 ppb
Scaling: Range high:............................................................ 1.00 ppm
Signal Output 2
Parameter: ......................................................................Temperature
Current loop:....................................................................... 4 –20 mA
Function:.................................................................................... linear
Scaling: Range low:.................................................................. 0.0 °C
Scaling: Range high:............................................................... 50.0 °C
Alarm Relay
Alarm Si 1:
Alarm high:........................................................................... 5.00 ppm
Alarm low:............................................................................... 0.0 ppb
Hysteresis: ............................................................................... 10 ppb
Delay:............................................................................................. 5 s
Sample Temp.: Alarm High: ....................................................... 55 °C
Sample Temp.: Alarm Low: .......................................................... 5 °C
Sample Flow: Flow Alarm: .............................................................yes
Sample Flow: Alarm High: ......................................................500 B/s
Sample Flow: Alarm Low: ...........................................................5 B/s
Case temp. high:........................................................................ 65 °C
Case temp. low:........................................................................... 0 °C