
A-96.250.671 / 070222
AMI Silica
Troubleshooting
70
7.4. Opening the peristaltic pump housing
For some electrical connections (e.g. when replacing suction lanc
-
es), it is necessary to open the housing of the peristaltic pump. To do
this, proceed as follows:
1
Switch off the analyzer according to
2
Remove the protection cap and all pump tubes as described in
.
3
Unscrew the 4 screws of the peristaltic pump housing and re
-
move the cover.
4
Disconnect the motor connector [A].
5
Feed the cable into the housing through one of the PG7 cable
glands.
6
Connect the cable to the terminal block of the peristaltic pump ac
-
7
Reassemble in reverse order.
A
Motor connector
A