A-96.250.671 / 070222
AMI Silica
Maintenance
57
6.8. Cleaning the solenoid valve
Disassemble
the solenoid
valve
The solenoid valve is mounted below the constant head. The sole
-
noid valve should be disassembled if it does not switch anymore or if
it is clogged.
1
Switch off the instrument according to instructions in
eration for Maintenance, p. 48
.
2
Loosen the nut (A).
3
Remove the solenoid coil (B) from
the valve body (C).
4
Loosen the fixing screws of the
valve body with a 2.5 mm allen key
(D).
A
B
C
D