AMI Codes-II TC
Program List and Explanations
A-96.250.631 / 280219
85
4.3.1
Log Interval:
Select a convenient log interval. Consult the table
below to estimate the max logging time. When the login buffer is
full, the oldest data record is erased to make room for the newest
one. (circular buffer)
4.3.2
Clear Logger:
If confirmed with
yes
, the complete logger data is
deleted. A new data series is started.
4.4 Display
Process values are displayed on two screens. Toggle screens with
the [
] key. Each screen displays max. 3 process values.
4.4.1
Screen 1
4.4.1.1
Row 1
4.4.1.2
Row 2
4.4.1.3
Row 3
Possible settings for all rows are:
None
Total chlorine 1
Total chlorine 2
Calc. dichloramine
pH
4.4.2
Screen 2
Same as screen 1.
5 Installation
5.1 Sensors
5.1.1
Dimension
: The measuring value can be displayed as ppm or mg/l
5.1.2
Ref. Verification:
Set absorbance value of verification kit according
to label.
Range: 0.200–0.600
Interval
1 s
5 s
1 min
5 min
10 min
30 min
1 h
Event
Driven
Time
25 min
2 h
25 h
5 d
10 d
31 d
62 d