A-96.250.881 / 070622
AMI Codes-II O3
Maintenance
51
6.6. Cleaning the Photometer
Clean the photometer after indication by alarm (E020, FOME dirty).
Switch off the instrument according to instructions in
.
Material
Small brush.
Procedure
1
Close the flow regulating valve [A].
2
Wait until the sample flow through the photometer has stopped.
3
Unscrew the screw plug [B] from the photometer [C].
4
Clean the Photometer with a small brush [D].
5
Screw the plug to the photometer.
6
Open the flow regulating valve.
A
B
C
D
Flow regulating valve
Screw plug
Photometer
Brush
A
B
C
D
C