SVAN 979 User Manual
2.
Set the calibration level (
Cal. Level
) appropriate to the used calibrator. The
default level for calibration in the vibration mode is 10 m/s
2
at 159.2 Hz.
Remember to change this level if using an alternative vibration calibration
signal source.
3.
Attach the instrument’s accelerometer to the vibration calibrator using an
appropriate or recommended fixing method.
4.
Switch on the calibrator and wait approximately 30 seconds before starting
the calibration measurement.
5.
Start the calibration measurement by pressing the
<Start>
key.
The instrument performs series of 1-second consecutive measurements
displaying levels of the measured calibration signal.
If maximal difference between three 1-second consecutive results (RMS(HP10))
is less than
0.05dB
, the calibration measurement will be stopped, and the
calibration factor calculated. The measurement can be always stopped by the
<Stop>
key.
The
Calibration Result
(measured reference signal without calibration factor
correction) and the
New Calibration Factor
(difference between the Calibration
Level and the Calibration Result, calculated in dB) are displayed.
6.
Press
<ENTER>
to save the new calibration factor (
Enter to Confirm
), or
press
<ESC>
to reject it.
7.
Press
<ENTER>
to exits the
Calibration by Measuement
screen.
It is recommended to repeat the calibration measurement few times. Obtained results should be almost the
same (with
0.1 dB difference). Reasons for unstable results are as follows:
•
the accelerometer is not properly attached to the calibrator,
•
there are external disturbances,
•
the calibrator or the measurement channel (the accelerometer or the instrument itself) are damaged.
8.
Detach the accelerometer from the calibrator.
Note: During the calibration measurement, the external disturbances (vibrations or acoustic noise)
should not exceed a value of 1/10 of the level of the calibration level signal being used.
3.3.6
History of performed calibrations
– Last Calibration
The
Calibration History
screen displays
records of performed calibrations.
To review the calibration records, select the
required line in the
Calibration History
screen and press
<ENTER>
.
<ENT>
Summary of Contents for SVAN 979
Page 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Page 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Page 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Page 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Page 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Page 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Page 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Page 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Page 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Page 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Page 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...