19
thin-walled glass.
•
Glass <1.5 mm O.D.
•
Transfer of model P-87 puller programs to P-97
Use the DELAY MODE for:
•
Patch programs using thick-walled glass
•
Single and multiple cycle programs
•
Glass >1.2mm O.D.
•
Thick-walled glass
•
Transfer of model P80-PC puller programs to P-97.
3.3.4
AIR TIME AT START OF PULL < 4 >
Controls the length of time the humidity control chamber is purged with dehumidified air
prior to execution of the pull sequence. The intent of the control is to minimize and stabilize
the chamber humidity during the pull sequence. Units are in seconds.
Range
1 to 255 seconds
Default
5 seconds
Recommendations
For humid environments
20 to 30 seconds
3.3.5
AIR TIME AT END OF PULL < 5 >
Controls the length of time the chamber is cooled after completion of the pull and the length
of time the end of pull information is displayed. Units are in seconds. The time you enter is
actually an offset added to an automatic 3-second chamber cooling cycle [e.g., A setting of 5
would result with 8 seconds of chamber cooling].
Range
0 to 255 seconds
Default
5 seconds
Recommendations
For single cycle programs
5 seconds
For multicycle programs
(to allow additional cooling of the
heating clamp assembly)
25 seconds
3.3.6
RESET TIME AND DATE < 6 >
This function allows you to set the system clock. The clock is used to establish the date and
time of the last program edit as displayed on the program header line.
3.3.7
WRITE-PROTECT THE PROGRAM < 7 >
Activate (lock) or inactivate (unlock) the WRITE PROTECTION mode. Unlocked mode (0)
allows program to be edited. Locked mode (1) prevents program from being edited. The
write-protect status is reported on the program header line.
P-97 FLAMING/BROWN MICROPIPETTE PULLER OPERATION MANUAL – REV. 2.43 - DOM (20161118)