SUPMEA SUP-LDG User Manual Download Page 1

User's

 

Manual

U-

PDCLDG

-MYEN

3

Electromagnetic

 flow meter

Supmea Automation Co.,Ltd

.

www.supmea.com

Headquarters

5th floor,Building 4,Singapore Hangzhou Science Technology Park,No. 6 street,
Hangzhou Economic Development Area,Hangzhou 310018,China

Singapore 

2 Venture Drive #11-30 Vision Exchange Singapore

Philippines

Majestic Subdivision, Lot 1, 1800 Rainbow St, Marikina, 1811 Metro Manila, Philippines

[email protected]

Summary of Contents for SUP-LDG

Page 1: ...ers 5th floor Building 4 Singapore Hangzhou Science Technology Park No 6 street Hangzhou Economic Development Area Hangzhou 310018 China Singapore 2 Venture Drive 11 30 Vision Exchange Singapore Phili...

Page 2: ...sses caused by false operation Note Modification of this manual s contents will not be notified as a result of some factors such as function upgrading We try our best to guarantee that the manual cont...

Page 3: ...6 Nameplate 11 Chapter 3 Installation 12 3 1 Installation tips 12 3 2 Storage 12 3 3 Pipeline design 12 3 4 Pipe design 12 3 5 Installation conditions 15 3 6 Mechanical installation 19 3 6 1 Installat...

Page 4: ...51 7 2 Pulse Frequency Current Output 52 7 2 1 Pulse equivalent output 52 7 2 2 Frequency Output 53 7 2 3 Current Output 53 7 3 Communication 53 7 3 1 Registered Address 54 7 3 2 Communication Configu...

Page 5: ......

Page 6: ...ind you data transmission on the Internet such as communicating via email may possibly meet security vulnerabilities We can t give security guarantee that data will definitely not be obtained by a thi...

Page 7: ...the corresponding standard sales terms applies as well and the clause is the basis of purchase contract 1 1 4 Document details In order to avoid harm or damage to the equipment when used improperly pl...

Page 8: ...d may damage the equipment itself or the operating factory facilities Note Such warnings must be paid attention to Any slight negligence may also lead to functional fault of the equipment itself 1 2 S...

Page 9: ...ote Please check the packing list to make sure that all the goods you received are integrated Note Please check the nameplate of the equipment and confirm whether the delivered contents are consistent...

Page 10: ...Introduction www supmea com 5 3 Signal cable 4 User manual 2 Compact type flowmeter DN65 refer to type selection manual for specific parameters Figure 2 1 Compact type electromagnetic flowmeter 2 Use...

Page 11: ...ction density of the magnetic field and the width of the conductor the inner diameter of the flowmeter measuring tube and the flow rate of the medium can be obtained by calculation The induced electro...

Page 12: ...table excitation current for the sensor meanwhile amplify the induced electromotive force obtained by the sensor and convert it to standard electrical signals or frequency signals at the same time it...

Page 13: ...c field 7 Casing Protect and seal the meter 2 4 Application introduction Electromagnetic flowmeter applies only to measure the real time flow rate of an electrically conductive liquid or liquid solid...

Page 14: ...uction 1 Remote type Figure 5 L N 100 240VAC power supply Ground ALM1 ALM2 Relay out POUT PCOM Pulse Frequency output 485A 485B RS485 communication IOUT ICOM 4 20mA output EXT1 EXT2 Excitation signal...

Page 15: ...com 10 2 Compact type Figure 6 L N 100 240VAC power supply 485A 485B RS485 communication IOUT ICOM 4 20mA output connection POUT PCOM Pulse Frequency Relay out CCOM RS485 communication ground Converte...

Page 16: ...eters of the meter are preset according to the requirements of the order thus users are not required to set the parameters prior to operation Instead you need to check whether the parameters indicated...

Page 17: ...our order Check whether the power supply indicated on the nameplate is correct If not correct please contact the manufacturer 3 2 Storage 1 The instrument shall be stored in a dry and clean place 2 Av...

Page 18: ...bration In this case brackets for fixing pipes on both sides of electromagnetic flowmeter shall be provided 6 Part of the sensor of electromagnetic flowmeters with IP68 3 m under water protection leve...

Page 19: ...gure 10 these devices can ensure the continuous operation of equipment system when the flowmeter is out of service 6 Support of electromagnetic flowmeter Do not install the electromagnetic flowmeter o...

Page 20: ...he meter the user shall make the flow arrow consistent with the on site process flow Figure 11 shows the preferred location for installing the electromagnetic flowmeter Figure 11 2 Installation direct...

Page 21: ...d with pipes Pipes shall be arranged to ensure that the electromagnetic flowmeter measuring tube is always filled with liquids Figure 12 For liquids or slurries containing solid particles it is sugges...

Page 22: ...valves are generally installed on the downstream of the electromagnetic flowmeter Figure 14 6 For pipes with open discharges the electromagnetic flowmeter shall be installed at the bottom section low...

Page 23: ...ry to mount the flowmeter on the valve s upstream side for preventing a decrease of pressure inside the pipe possibly consequently avoiding the possibility of air bubbles ensure that no gas can be sep...

Page 24: ...erference through the grounding of flowmeter casing thereby improving measurement accuracy 3 6 Mechanical installation 3 6 1 Installation of flowmeter pipeline 1 Prior to installation the pipeline sha...

Page 25: ...y the flange surface which cannot be stressed otherwise the lining may be damaged to disable operation of the meter 3 Protection of flange surface After unpacking pay attention to protect the flange D...

Page 26: ...please take the following measures A Check the tightness of the covers and the wiring terminals to ensure that no moisture and water enters into the meter B Conduct regular inspection Check against th...

Page 27: ...l be taken to prevent the lining from being deformed due to heat 4 If it is installed in a well or immersed in water apply sealant on the terminal box of the sensor after the system is installed and d...

Page 28: ...5 150 322 82 95 65 4 14 20 150 322 78 105 75 4 14 25 150 312 78 115 85 4 14 32 150 327 74 135 100 4 18 40 150 335 74 145 110 4 18 50 200 354 86 160 125 4 18 65 200 366 92 180 145 8 18 80 200 385 92 19...

Page 29: ...0 20 25 600 600 940 408 780 725 20 30 700 700 1048 520 895 840 24 30 800 800 1160 580 1010 950 24 34 900 900 1260 660 1110 1050 28 34 1000 1000 1370 720 1220 1160 28 34 1200 1200 1585 1130 1405 1340 3...

Page 30: ...ctrical equipment Tips Please check the nameplate of the equipment and confirm whether the delivered contents are consistent with your order and check whether the voltage indicated on the nameplate is...

Page 31: ...to operate on the electrical equipment 1 Compact type caliber 65 refer to type selection manual for specific parameters Figure 20 Connection description 1 Excitation output EXT1 Sensor excitation coil...

Page 32: ...akage current may cause serial mode interference to the measurement signal or in a serious case the electromagnetic flowmeter cannot work 1 The measurement sensor must be correctly grounded 2 The grou...

Page 33: ...240VAC 50Hz 60Hz 1 L AC live line 2 N AC neutral line 3 Connect ground wire to the ground screw 2 24VDC power supply Figure 22 Tips Allowable range 22VDC 26VDC 1 24 24VDC Power supply positive pole 2...

Page 34: ...so as to make sure that you use the equipment in a safe and effective way Figure 23 Currentoutput 1 IOUT ICOM 4 20mAoutputconnection Communicationoutput 2 485A 485B RS485communicationoutput 3 CCOM RS...

Page 35: ...ratio 1 1 Fmax 5000cp s Electricwiringdiagram750 under load Imax 22mA Current flow percent 6 Electricwiringdiagram Figure 24 Additional remarks pulse output is OC gate output it needs external power s...

Page 36: ...connection in the power supply is correct 4 Tighten the converter shell back cover 5 Tighten the back cover of the converter housing 5 2 Converter start up The measuring instrument is consisted of mea...

Page 37: ...System alarm information 6 Cumulative amount and so on Display information Positive flow accumulation Negative flow accumulation Net flow accumulation V current flow rate MT Current conductivity 7 Op...

Page 38: ...peration use a finger to click on the icon for more than half a second and release to finish button operation for once Except key combination it is forbidden to put other fingers on the other photoele...

Page 39: ...ss mechanical keys to enter configuration mode is shown in the next chapter Figure 28 6 4 Quick setup menu Key parameters to facilitate the manufacturer and user to quickly set up the meter Press on a...

Page 40: ...1 2 density parameter 1 2 Fluid density Figure User 0 000 99 000 1 000 Used to calculate the mass flow QM VM When the flow unit is volume this parameter are not displayed Density unit g cm3 1 3 Time c...

Page 41: ...or permission Option User Y N N Indicate whether to enable peak inhibition function used for filtering interference signals When it s set to be N 1 8 1 9 configuration screens do not display When the...

Page 42: ...the revised flow velocity is called modified velocity the modified computation formula is as follows At the interval of the modified point 1 The original flow velocity The modified point 2 The modifie...

Page 43: ...point 2 when The flow rate function is disabled this parameter does not display 1 14 Flow correction coefficient 2 Figure Manufacture 0 0 99 999 1 000 Flow correction factor 2 This parameter is not di...

Page 44: ...0 99999 0 000 Used for adjusting the output current value I Kx B 2 3 Output current Display User 4 00 20 00 Display the current value mA of the current output 3 Pulse frequency alarm output 3 0 Pulse...

Page 45: ...on integer Figure Manufacturer 0 999999999 0 Set reverse total integer part 4 5 Negative accumulation decimal Figure Manufacturer 0 0 0 999 0 0 Set reverse total decimal part 5 Alarm contacts 3 0 set...

Page 46: ...ower limit alarm output switch When the real time flow is smaller than the flow rate lower limit value contact 1 outputs alarm signal automatically See 7 2 for details When allowed alarm output config...

Page 47: ...crystal display 8 3 Modbus address Figure User 1 247 8 Communication Protocol instrument address based on the RS485 protocol Modbus RTU 8 4 Baud rate Option User 1200 2400 4800 9600 19200 38400 57600...

Page 48: ...uring line it is recommended to use double shielded wire when the wiring distance is 20m otherwise it will affect empty detection function 9 2 Empty pipe check permission Option Manufacturer Y N Y Set...

Page 49: ...er the circumstance of sensor symmetry and wiring is good well shielded and within the scope of code value 0 1 no need to adjust 10 4 Sensor coefficient Figure Manufacturer 0 99999 The flowmeter coeff...

Page 50: ...Y N N Set to Y to make the test flow rate effective and automatically return to N after power off 11 1 Flow rate Option Manufacturer 99 999 99 9 99 1 000 To set the simulated flow rate it will take ef...

Page 51: ...er password 100000 used for modifying the manufacture level parameter The initial manufacturer password 300000 to set up parameter quickly After entering the configuration parameters the parameters ca...

Page 52: ...6 Operation www supmea com 47 6 6 2 Display measurement This picture will display after startup Forward cummulant Reverse cummulant Net cummulant V Current flow velocity MT Conductivity equivalent Fi...

Page 53: ...Chapter 6 Operation www supmea com 48 6 6 3 Flow setup and analog output menu Figure 31...

Page 54: ...Chapter 6 Operation www supmea com 49 6 6 4 Pulse output and total set menu Figure 32 6 6 5 Relay out setup menu Figure 33...

Page 55: ...Chapter 6 Operation www supmea com 50 6 6 6 System function empty pipe function sensor function test function setup menu Figure 34...

Page 56: ...stem Information Sheet Display Alarm content Mtsnsr Sensor empty pipe Hi The current real time flow rate exceeds the setting flow limit Lo The currentreal timeflowrateis belowthesettingflowlower limit...

Page 57: ...nter the above situation the decimal part of 2 5 pulses will automatically get into the next second output data loss will not happen The pulse equivalent shouldn t be set too small when the pipe flow...

Page 58: ...ther intelligent instruments such as digital display table recorder PLC DCS etc The current output type 4 20mA The current valve corresponds to real time flow rate 20mA corresponds to range limit 4 mA...

Page 59: ...lectric conductivity float 106 Forward flow accumulation of integer ulong 108 Forward flow accumulation of decimal ulong 110 The decimal part magnifies by 100 times 123 stands for 0 123 Reverse flow a...

Page 60: ...ication frame Example Long integer 16909060 01020304H 03 04 01 02 Floating number 4 00 40800000H 00 00 40 80 7 3 3 Readout Real time Quantity Floating point Communications Example Real time floating p...

Page 61: ...e with GB T9119 2000 standard carbon steel Optional stainless steel flanges other standard flange can be customized Pressure rating High pressure can be customized DN6 DN80 PN 4 0MPa DN100 DN150 PN 1...

Page 62: ...Maximum measurement error 0 1 Within the measuring range of temperature sensor Operating Environment Graphic display Monochrome LCD white backlight Size 128 64 pixels Display function 2 measurement v...

Page 63: ...n Max 15VA Signal cable Apply only to remote type Shielded cable Signal section wire 0 5mm2 Cu AWG20 Output Current output function Measurement of volume and quality in the case of constant density Se...

Page 64: ...that contains other oxidants For example Hypochlorite solution higher than room temperature is strongly corrosion resistant to sea water Not Applicable Reducing acid and chloride such as hydrochloric...

Page 65: ...0 300 530 150 6 36 60 80 400 500 763 200 11 3 100 120 600 800 1357 250 17 7 160 200 800 1000 2120 300 25 45 250 300 1200 1600 3054 350 34 6 300 400 1600 2000 4157 400 45 2 400 500 2000 2500 5429 450 5...

Page 66: ...5 4 42 8 84 17 67 22 09 44 18 441 8 530 16 662 7 DN150 6 36 12 7 25 5 31 81 63 62 636 2 763 44 954 3 DN200 11 3 22 6 45 2 45 55 113 1 1131 1357 2 1696 5 DN250 17 7 35 4 70 7 88 36 176 7 1767 2110 4 26...

Page 67: ...ers www supmea com 62 8 5 Accuracy Reference condition 1 Medium water 2 Temperature 20 3 Pressure 0 1MPa 4 Front straight conduit 5DN Rear straight conduit 2DN Figure 36 1 X m s Flow rate 2 Y Actual m...

Page 68: ...or the maintenance and repair of instruments 3 The measurement accuracy is not affected by changes in physical parameters such as temperature pressure density viscosity conductivity as long as the con...

Page 69: ...lanar electrode type is suitable for measuring the liquid flow rate containing other impurities in the medium 9 3 Main technical specification 1 Applicable pipe diameter 200 2000mm 2 Flow rate measure...

Page 70: ...t for the electrode tip or the inner wall of the measuring tube the other parts are insulated from the fluid to be measured Excitation system The excitation system is used to generate a working magnet...

Page 71: ...test and is used to connect with the mounting ball valve and insert the electromagnetic flowmeter sensor Sealing mechanism composed of pressing screw seat pressing nut rubber washer and set screw made...