31
4.12.5 Analogue output
★
Output channel: 01-06
★
Input channel: The input channel corresponding to the output channel. For example,
"Output 02" corresponds to "Channel 05", and the analog output of No.2 channel follows the
transmitted output of the measured value of input channel 5 .
★
Output type: It may select 0~10 mA, 0~20 mA, 4~20 mA, 0 ~ 5V, 1 ~ 5V, 0 ~ 10V, nil.
★
Output low Limit and output high limit: Adjust the high and low limit ranges of the
transmitted output. Set the range : -9999
~
19999. The bit number of decimal points in the channel
is determined by modifying the position of decimal points in the high limit of the range.
★
Kx+b: Linear correction formula for analog output, b: output zero point transfer volume, K:
output amplification ratio.
Operation: In this screen,
Move the cursor by pressing the keys "Left shift" and "Right shift".
Select the wanted contents by pressing the keys “Down shift", "Up shift", " Page down " and
" Page up".
Press the "Enter" key to confirm the execution;
Press the "Esc" key to exit this screen. If the content of the screen is modified, it will pop up
the dialogue box "Do you save the modified parameters?" In the dialog box, select "Yes" to save
the modified parameters and "No" to save the modified parameters.
Note: As channels7~12 terminals of the output channel and the alarming channel are shared.
For example, the instrument has 8 channels for alarming outputs and 3 channels for analog outputs.
Our company follows the principle of priority of alarming channels, so 01 and 02 of the output
channels are occupied by the alarming channels. The output 03 corresponds to No.1 channel
analog output, and the output 04 corresponds to No.2 channel analog output, and so on.