Chapter 2 Starting Measurement
www.supmea.com
14
(8)Press key
to enter M24 window to install the transducers on the pipe, and
then press
key to go to M01 for the results.
The first-time users may need some time to get familiar with the operation.
However, the user friendly interface of the instrument makes the operation quite
easy and simple. Before long, the user will configure the instrument with very little
key pressing, since the interface allows the user to go to the desired operation
directly without any extra steps.
The following tips will facilitate the operation of this instrument.
(1) when the window display is between M00 to M09, press a number key x , the
user will go directly to the M0x window. For example, if the current window displays
M0l, press 7 and the user will go to M07.
(2) when the window display is under M00 to M09, press the ENT key and the user
will go to M90; press ENT key to return. Press the dot key to go to M11 when the
window display is under M25, press ENT key to go to M01.
2.7 Transducers Mounting Allocation
The first step in the installation process is the selection of an optimum location in
order to obtain a more accurate measurement. For this to be completed
effectively,a basic knowledge about the piping and its plumbing system would be
advisable.
An optimum location would be defined as a straight pipe length full of liquid that is
to be measured. The piping can be in vertical or horizontal position. The following
table shows