Supertech agroline Superpro User Manual Download Page 10

Battery

The meter is supplied with a 9-volt alkaline battery. This should be installed in the compartment on the
underside of the meter.  When the battery is in need of replacement the message

Replace battery

will

appear on the display.
Only replace it with a similar high quality 9-volt alkaline battery. Removing the battery has no effect on the
information stored in the meter.
9-volt batteries should not be thrown away with trash. This may risk contact with other batteries or pieces
of metal which could lead to danger.
In case of battery leaking please remove carefully by using gloves. After removal please clean the device
carefully with a clean cloth.

Storage and Advice for General Usage

It is recommended that you store the meter at all times in the carrying case provided, so that the instrument
is not exposed to large temperature fluctuations and that it is kept free of moisture. If the instrument is
taken into a warmer environment from a cold storage location, it is recommended that you allow the unit to
acclimatise so that condensation does not affect the results of the measurement. At the end of the season,
wipe all surfaces with a well-wrung damp cloth and remove the battery from the meter. Clean all parts of
the grinder mechanism and lubricate the spindle with a thin layer of non-corrosive oil.

Auto Turn OFF time - Setting

Our Superpro is pre programmed to Turn-OFF after approximately 90 seconds of inactivity.
If the device turns off by itself while using it please turn on the meter by pressing ON/TEST.

How to make your own grain calibration

If you need to make a calibration scale for a particular crop that is not included in the meter, this can be
done as follows:

1. Turn on the meter by pressing ON/TEST.
2. When MEASURE appears press ON/TEST again.
3. Use the arrow keys

or

scroll through the crop options until Bit is displayed.

4. Fill the test cell with a sample of the desired crop for which you know the moisture content, grind

the sample and perform the test as described earlier.

5. Note down the value which appears in the display, eg. 03.65, and the temperature, eg. 19C, together

with the moisture content in the crop. Perform this process repeatedly using samples with different
levels of known moisture content until the calibration scale meets the range and requirements
needed. The more samples that are used the more accurate the calibration will be. Try always to use
samples at the same temperature.

6. Plot the values on a graph, either manually or using Excel, with moisture content on the y-axis and

Bit on the x-axis and draw a line or curve that best fits the points. The calibration is now completed

Ver 1.1.0

GB

Page 10/12

Summary of Contents for Superpro

Page 1: ...Superpro User Manual Instruction manual Bedienungsanleitung Notice d utilisation Manuale di Istruzioni Gebruiksaanwijzing Bruksanvisning Instrucciones de uso Instrukcja obs ugi Ver 1 1 0 GB Page 1 13...

Page 2: ...the Meter 8 Superpro Settings and Menu s 9 MEASURE 9 BATTERY 9 CLEAR AVERAGE 9 CALIBRATION 9 Some more valuable information about the moisture meter 9 Using the averaging facility 9 Battery 10 Storag...

Page 3: ...omplies with EU Directives and standards among other regulations for electrical safety and electromagnetic compatibility Inaccurate usage or handling may however result in damage to the product and sm...

Page 4: ...of ABS plastic Supplied in shockproof carrying case Introduction Thank you for choosing Supertech s Superpro Grain Moisture Analyser This moisture meter is a ground grain meter with built in grinder w...

Page 5: ...9 and 11ml measuring cups Battery Operator s manual Brush for cleaning grinder Shockproof carrying case Part Locator 1 Grinder handle with 9 11ml measuring cups 2 Grinder housing lower grinder cell 3...

Page 6: ...TEST button Select the crop to be tested by using the arrow keys or To select a different crop press the ESC OFF button to return to MEASURE and use arrow keys or to cycle through the options 3 Fill t...

Page 7: ...nt Remnants of sticky or moist material can be removed by grinding a dry grain sample Note When grinding crops with high water content there may be a risk of the two grinder discs touching one another...

Page 8: ...ing grain from the same sample in both cases and adjusted as required Each crop calibration in the meter can if necessary be adjusted individually up or down by up to 3 9 in increments of 0 1 This is...

Page 9: ...pear in the display followed by an automatic return to MEASURE CALIBRATION The calibration menu is used to recalibrate the meter Differing growth factors between years and between varieties can have a...

Page 10: ...the grinder mechanism and lubricate the spindle with a thin layer of non corrosive oil Auto Turn OFF time Setting Our Superpro is pre programmed to Turn OFF after approximately 90 seconds of inactivit...

Page 11: ...ications Measuring range Moisture 5 to 50 Depending on crop type Operating range Temperature 0 to 50 C Resolution 0 1 moisture Temperature compensation Automatic Weight including case 2 kg Dimension 1...

Page 12: ...Failing to meet this requirement can result in heavy metals accumulating in the environment and potentially be damaging towards humans animals and natural resources EU Declaration of Conformity Accor...

Reviews: