Calibration Coefficients:
A calibration printout with the coefficient values can be obtained and kept in a log which is useful for
documentation and tracking any drift trends in the sensors or changes in the channel definitions.
Performing a current value calibration creates a new coefficient number for that channel or channels. All
other channel coefficients should remain the same. The offset value changes every time the system is
zeroed.
1. From the main
WinDyn
menu, select
Tools>>View Sensor Calibration
.
2.
Using the Tutorial at the bottom of the window, follow the steps to open the file and print the calibration
values.
6.4 2640 Circuit Board LEDs
1. Both LEDs should be ON when powered up.
2. Active LED should be blinking when powered
up.
3. Both LEDs should be ON when connected to
the computer network. The top LED will be
flashing when WinDyn or NetDyn are active.
Figure 6.1 2640 Circuit Board LEDs
#1 LEDs
#2 LED
#3 LEDs
6.0 Maintenance