17. Maintaining and Cleaning the Oscar 2
The Oscar 2 system is designed to perform in conformity with the description contained in this user manual
and accompanying labels and inserts, when assembled, operated, maintained, and repaired in accordance
with the instructions provided. After use, it is important to perform preventative maintenance to ensure the
safe and efficient operation of the monitor. It is your responsibility to:
•
Check calibration of the device every two years.
•
Never knowingly use a defective device.
•
Immediately replace parts that are broken, worn, missing, incomplete, damaged or contaminated.
•
Contact the nearest authorized service center should repair or replacement become necessary.
Further, the user of the device bears sole responsibility for any malfunction that results from improper use,
faulty maintenance, improper repair, damage or alteration by anyone other than SunTech Medical or autho
-
rized service personnel.
While usage will have an impact, it is expected the monitor will be in service for 7 years. Typically, an electro
-
mechanical pump determines the lifetime of the monitor. Service and support, including relevant accessories,
will be available up to 7 years following the last date this product is manufactured by SunTech Medical. For a
list of service centers, see Section .
Cleaning After Use
Oscar 2
The Oscar 2 monitor cannot be sterilized. DO NOT immerse the monitor in any fluid, or attempt to clean with
any liquid detergents, cleaning agents, or solvents. The following methods and cleaning agents have been
tested and confirmed not to affect the device housing:
•
Clorox Wipes
•
Dispatch Wipes
•
Quternary Disinfectant Cleaner
•
Sani-Wipes
Cleaning Method
1.
Turn device off or remove batteries and close battery compartment.
2.
Wipe down device with cleaning agent
3.
Allow device to sit until dry.
Page 69 of 79