background image

13 

Patient Care Functions 

Placing the Patient on the Mattress Surface 

Place the patient on the mattress surface from a bed or stretcher with a transfer device.  The 
mattress should be set in the Auto Firm mode.

 

In order to ensure proper immersion and 

envelopment of the patient, the user should:  

1. Position patient on surface in center of bed.
2. Initialize soft/firm settings on the Control Unit/Blower.
3. Wait a moment to allow internal sensors to activate pressure redistribution. Generally,

depending on patient body makeup, initial pressure redistribution is complete in
approximately 2-3 minutes.

4. Elevate the head of the bed to at least 30 degrees.
5. Unzip the mattress and visually inspect the height of the cells for sufficient inflation.

There should be several inches between the patient and the frame of the bed.  Adjust the
air flow accordingly.  Ask the patient if they can feel the bed frame beneath them. If yes,
add air incrementally. Repeat until patient no longer feels the frame beneath them.

6. If the patient cannot reply verbally, unzip the mattress and visually inspect the height of

the cells for sufficient air inflation.  There should be several inches between the patient
and the frame of the bed.  Adjust the air flow accordingly.

7. CPR: The standards for life support recommended by the American Heart Association for

performing  CardioPulmonary  Resuscitation  (CPR)  recommend  a  hard  level  surface  for
performing CPR, moving the person to the floor if possible. For performing CPR, place
the CPR board, lower the head of the bed, position the patient on their back and follow
standard CPR procedures of the facility.

In some cases, the mattress may include an auto-fowler feature.  This means that when the head 
of the bed is greater than 30 degrees, the mattress will automatically increase the amount of air to 
the sacral area.

 

Positioning the Patient 

When moving a patient the Control Unit/Blower should be in the Auto Firm Mode. 

To reposition the patient, change the Control Unit/Blower to Auto Firm Mode.  This makes the 
mattress  surface firm  and facilitates the repositioning of the patient  with  less strain  on the care 
provider.    When  the  patient  has  been  repositioned,  press  auto  firm  to  return  to  the  previous 
setting. 

NOTE: 

 DO NOT leave a patient unattended on the mattress surface with the safety side rails in 

the down position.  When leaving a patient, secure the safety side rails in the up position.  Make 
sure the safety side rails are high enough to properly protect the patient when the mattress is fully 
inflated, while continuing to be mindful of the FDA guidelines on bed rail entrapment. 

Backrest Up or Fowler Position 

When  the  patient’s  backrest  is  elevated,  it  may  be  necessary  to  manually  increase  the  mattress 
firmness  to  compensate  for  the  additional  weight  placed  in  the  center  portion  of  the  mattress.  

Summary of Contents for Static Sapphire Pulse

Page 1: ...User Manual 8083 Rev 8 0 03 04 2021 User Manual Sapphire Series Static Low Air Loss Mattress Replacement System...

Page 2: ...rence 12 Patient Care Functions 13 Placing the Patient on the Mattress Surface 13 Positioning the Patient 13 Backrest Up or Fowler Position 13 Prone Position 14 Bedpan Placement Removal 14 Removing th...

Page 3: ...2 Warranty Provisions 32 User Assistance Information 33 Warranty Information 32 Return Exchange Goods Policy 32...

Page 4: ...he user to the presence of important operating maintenance or servicing instructions Disregarding a warning could result in patient and or user injury as well as damage to equipment Electrical Shock H...

Page 5: ...perly protected from electrical shock This symbol marks the location of the leakage test point screw The hazards and warnings are indicated on the shipping container by this label This symbol marks th...

Page 6: ...damage or failure Monitor the unit to ensure that it functions in the proper operating temperature CAUTION Keep out of direct sunlight CAUTION DO NOT store in temperatures below 0 F 18 C or above 95...

Page 7: ...t the foot of the bed or sits on the floor under the bed The Sapphire Series Static Low Air Loss Mattress Replacement System has a detachable hospital grade electrical cord and a control panel with se...

Page 8: ...d Frequency 220 Volts Input Power 50Hz 230 Volts Degree of Shock Protection Type B Maximum Relative Humidity 95 Storage Temperature 18 C to 35 C 0 F to 95 F Environmental Conditions Product must be st...

Page 9: ...ght company and file a damage complaint immediately Environmental Conditions CAUTION Keep out of direct sunlight DO NOT expose to temperatures greater than 35 C 95 F or below 18 C 0 F DO NOT expose th...

Page 10: ...attress with the Sapphire Series Static Low Air Loss Mattress Replacement System Secure air mattress to the bed frame with straps provided Warning or Safety Instructions Relating to Set Up WARNING 110...

Page 11: ...d is unobstructed 6 NOTE Plug the 110V unit ONLY into grounded 110V AC outlet Plug the 220V unit ONLY into grounded 220V AC outlet 7 Turn master power switch ON located on the side of the unit 8 Press...

Page 12: ...utes if there are no changes to the blower settings the lockout feature will activate b Press and hold the Lockout button for 3 seconds to disengage the lockout feature 5 Quick Deflation or CPR Use a...

Page 13: ...3 sec Alarm Silence Mutes the audible alarm Visual light will not turn off until failure is resolved Autofirm Quickly inflates mattress to maximum firmness Fowler Used when head section of bed is elev...

Page 14: ...art Association for performing CardioPulmonary Resuscitation CPR recommend a hard level surface for performing CPR moving the person to the floor if possible For performing CPR place the CPR board low...

Page 15: ...ay be adjusted to increase the firmness of the center section after the pan is placed in position When the bedpan is to be removed logroll the patient off the bedpan and remove it Readjust the firmnes...

Page 16: ...n the manufacturer s instructions All cleaning solutions must be diluted in accordance with manufacturer s instructions All covers must be rinsed after every cleaning cycle Rinsing of the support surf...

Page 17: ...always be used as directed by the disinfectant s Safety Data Sheet 1 Prepare the disinfectant according to the manufacturer s recommendations 2 Prepare a separate bucket of warm fresh water to be used...

Page 18: ...used for rinsing the equipment after cleaning disinfecting procedures are completed as instructed 3 Allow the air cells to remain wet with disinfectant solution for the manufacturer s recommended con...

Page 19: ...ons to remove organic material and visible soil 6 Allow the control blower unit to remain wet with disinfectant solution for the manufacturer s recommended contact time 7 Rinse all surfaces of the con...

Page 20: ...ain and service this product Inspect top cover for punctures rips tears or damage Inspect mattress base for punctures rips tears or damage Connect the control unit and verify proper operation if insta...

Page 21: ...nction with the manufacturer supplied foam filter DO NOT attempt to install HEPA filter without the original equipment foam filter NOTE Ensure factory open cell foam filter is completely dry before pl...

Page 22: ...oting Storage If you store your product for more than 30 days we recommend Remove HEPA filter Place HEPA filter in an air tight plastic bag or plastic wrap ensuring filter is sealed to prevent exposur...

Page 23: ...flat on a clean level surface Avoid storage of other equipment on top of the support surface DO NOT expose the blower unit to humidity greater than 95 End of life Sunflower Medical L L C products must...

Page 24: ...ury as well as damage to equipment or other property The use of cables or accessories other than those for which the blower was designed or tested can significantly degrade emissions and immunity perf...

Page 25: ...gnetic radiation over the frequency range 80 MHz to 2 5 GHz The test was conducted on this blower and passed the requirements of IEC 61000 4 3 All pins of connectors have passed ESD testing List of Ca...

Page 26: ...ent Guidance RF Emissions CISPR 11 Group 2 The blower must emit electromagnetic energy in order to perform its intended function Nearby electronic equipment may be affected RF Emissions CISPR 11 Class...

Page 27: ...ality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1kV Differential 2kV Common A Mains power quality should be that of a typical commercial or hospital environmen...

Page 28: ...equipment should be separated from the blower by no less than the distances calculated listed below D 3 5 V1 Sqrt P D 3 5 V1 Sqrt P 80 to 800 MHz Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz V...

Page 29: ...olled The customer or user of the blower can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications Equipment and the blower as recom...

Page 30: ...l entrapment as serious injury can result These guidelines can be found on the FDA website When using an Air Therapy system the care provider is responsible for ensuring the mattress properly fits the...

Page 31: ...t punctured kinked or disconnected Ensure air cells are not punctured Check for proper connections from the hoses to the blower Make sure they are secure Check or clean air filters Ensure the CPR valv...

Page 32: ...l units Part Number 27000445 Connector Connects from mattress to blower control unit side panel Allows air movement to inflate mattress Part Number 27400035 Part Number 27400049 Part Number 27400050 H...

Page 33: ...Sunflower Medical LLC Warrants that all products provided under this agreement are free from defects in material and workmanship for the following stated warranty period from the date of delivery Sun...

Page 34: ...ser Assistance Information For questions or assistance with this product contact Sunflower Medical L L C at Sunflower Medical L L C 206 Jefferson Street PO Box 276 Ellis Kansas 67637 Phone 1 888 321 3...

Page 35: ...medical com Sunflower Medical L L C and Sunflower Medical L L C Rentals are trademarks of Sunflower Medical L L C Rentals L L C All specifications equipment and prices are subject to change without no...

Reviews: