Chapter 8: Maintenance
6. Press
ENTER
to confirm setting the install date to the current date and return to the previous
menu.
7. Press
EXIT
three times to return to the
Main Menu
.
Setting the Reagent Column and Zero Column Warning Levels
Set this warning level to specify the number of days to be alerted prior to the reagent column or zero
column needing to be replaced (1 through 365 days; default is 30). The warning message is displayed as a
"D" (for the reagent DI column) or a "Z" (for the zero column) and is displayed in the maintenance alert
area of the screen.
To edit the number of days:
1. Press
SETUP
from the
Main Menu
.
2. Highlight
MAINTENANCE
and press
ENTER
.
3. Highlight the appropriate menu option (
DI RESIN
for the reagent column or
ZERO BORON RES
for the zero column) and press
ENTER
.
4. Highlight
EDIT WARNING LEVEL
and press
ENTER
.
5. Use the
UP
and
DOWN
arrow keys to adjust the warning level to the desired number of days.
Once the warning level is set, press
ENTER
to return to the previous menu.
6. Press
EXIT
three times to return to the
Main Menu
.
Pump Tubing
.
The tubing for the peristaltic sample pump loses its elasticity over time and must be replaced yearly to
ensure proper flow through the Analyzer. The tubing must be purchased from SUEZ Customer Care or
your local service provider. The use of tubing from other sources or failure to replace the tubing on the
prescribed replacement schedule may degrade the quality of the Analyzer's measurement. Instructions on
how to replace the sample pump tubing are provided within the firmware.
Warning
Extreme caution must be exercised when operating the Analyzer with power
applied and the door open. Only qualified technicians should operate the
equipment in this manner.
Summary of Contents for Sievers UPW Boron Analyzer
Page 4: ...SUEZ 2018 4 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 16: ...SUEZ 2018 16 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 103: ...Notes SUEZ 2018 103 of 104 DLM 71000 03 EN Rev B 104...
Page 104: ...Notes SUEZ 2018 104 of 104 DLM 71000 03 EN Rev B 104...