Chapter 6: Setting Up Inputs and Outputs
Isolated Binary Outputs
There are 10 choices available for a binary output. Use the
UP
and
DOWN
arrow keys to scroll through
them.
OFF
The
OFF
setting disables the binary output and is the default setting for all four of the binary outputs.
BORON STREAM X
The
BORON STREAM X
setting causes the binary output to activate when a preset threshold is met.
STATE
The
STATE
setting will activate the binary output when the Analyzer is stopped (either by normal means or
a fatal error).
RES ERROR ANY STREAM
The
RES ERROR ANY STREAM
setting causes the binary output to activate when the resistivity is low
enough to trigger an error or warning.
ERROR
The
ERROR
setting causes the binary output to activate when an error condition occurs on the Analyzer.
RES ERROR STREAM X
The
RES ERROR STREAM X
setting causes the binary output to activate when the resistivity is low
enough to trigger an error or warning.
READY STREAM X
The
READY STREAM X
setting
causes the binary output to activate when a boron measurement has
completed on the specified stream.
RES WARN ANY STREAM
The
RES WARN ANY STREAM
setting causes the binary output to activate when the resistivity is low
enough to trigger an error or warning.
BORON ANY STREAM
The
BORON ANY STREAM
setting causes the binary output to activate when a preset threshold is met.
RES WARN STREAM X
The
RES WARN STREAM X
setting causes the binary output to activate when the resistivity is low enough
to trigger an error or warning.
Summary of Contents for Sievers UPW Boron Analyzer
Page 4: ...SUEZ 2018 4 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 16: ...SUEZ 2018 16 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 103: ...Notes SUEZ 2018 103 of 104 DLM 71000 03 EN Rev B 104...
Page 104: ...Notes SUEZ 2018 104 of 104 DLM 71000 03 EN Rev B 104...