Chapter 6: Setting Up Inputs and Outputs
Step Five: Configure Serial Communication Program for Retrieval of
Data from Data Card
Serial communication configuration can be accomplished by ensuring the following criteria are met:
• Match Baud Rates of the Analyzer and the computer (default setting = 19200)
• Set Data Bits on the computer to 8
• Set the Parity on the computer to
NONE
• Set the Stop Bits on the computer to 1
• Set the Flow Control on the computer to
NONE
For example, using HyperTerminal:
1. Under
PORT PROPERTIES
, click on the
SETTINGS
tab.
2. Click on the
ASCII Setup
button.
3. Click on the box next to "Echo Characters Locally".
4. Click the
OK
button twice.
Once the setup is complete:
1. Select
TRANSFER
on the toolbar.
2. Click on
CAPTURE TEXT
and input a file name to have the data stored in a text file.
3. Click on
START
.
Data collected over the RS-232 serial port will be transmitted in a comma-delimited format for incorporation
into various standard spreadsheet applications. Table 5 displays the types of data, field positions in which
the data appears in the output, and the output data format.
Table 5: RS-232 Data Format
Summary of Contents for Sievers UPW Boron Analyzer
Page 4: ...SUEZ 2018 4 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 16: ...SUEZ 2018 16 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 103: ...Notes SUEZ 2018 103 of 104 DLM 71000 03 EN Rev B 104...
Page 104: ...Notes SUEZ 2018 104 of 104 DLM 71000 03 EN Rev B 104...