Chapter 3: Installation
Figure 1: UPW Boron Analyzer
Sample Water Considerations
•
Sample water resistivity greater than 15 MΩ-cm
•
Minimal water particulates (15-micron filters provided)
•
Flow rate between 0.05 and 0.2 L/min. per sample stream
•
Water pressure between 20 and 100 psi at the Analyzer
•
Maximum sample tubing length from water sampling point less than 100 feet
•
Sample point(s) within 50 feet of tubing run from Analyzer preferred
.
Note:
There is an additional pressure drop of up to 2.5 psi and sample delay of up to
3 minutes for each additional 50 feet of sample line tubing.