![Suez Sievers M500e Operation And Maintenance Manual Download Page 136](http://html1.mh-extra.com/html/suez/sievers-m500e/sievers-m500e_operation-and-maintenance-manual_1389721136.webp)
C
ONFIGURING
R
ESULT
P
REFERENCES
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
136
SUEZ © 2022
3.
Navigate to the
Analyzer Settings
panel (
Settings
s
screen >
Instrument
Settings
tab).
4.
In the
Warning/Error Notification
area, select one of the following options:
•
Enable All Warning/Errors
— Select this option to display all warnings and
error notifications.
•
Disable Consumables Warnings
— Select this option to disable the
screen notification of an expiring/expired UV Lamp, Pump Head/Tubing, or
Resin Cartridge.
•
Disable All Warnings
— Select this option to disable the screen notification
of all warnings (no dialog box message, no yellow triangle at the top of the
screen, and no warning text on the lower-right screen). Error History
records, alarms, and analog outputs are independent of this setting and will
still occur.
•
Disable All Warnings and Errors
— Select this option to disable all
warnings and error notifications, including Error History records, alarms, and
analog outputs.
Configuring Result Preferences
Use the
Result Preferences
panel to select the results data, from online analysis
and grab sampling, to appear on the
Home
h
screen and
Data
D
screen (data
history). The panel also includes
Trend Graph Display
options.
Also, use this panel to set the Analyzer to automatically print results specific to
Grab
Protocols, System Protocols or
Online
modes. The frequency can be set to
every hour, every 2 hours, every 4 hours, or every 8 hours.
To configure result preferences
1. Navigate to the
Configuring Result Preferences
panel (
Settings
s
screen >
System Setup
tab).
2.
In the
Rejected Repetitions
area, select or de-select whether to include
rejected repetitions in one or more of the following:
•
View (on the Analyzer)
•
•
Export
3.
In the
Result Display Options (Enabled Columns)
area, select or de-select
which of the following columns to include:
•
TOC
•
IC