Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
135
SUEZ © 2022
To configure the display format settings
Configure the settings for the date format, the clock format, and number format for
display.
1. On the
Settings
s
screen, select the
System Setup
tab.
2.
On the
Local Setting
panel, under
Display Formats
, select one of the following
format options from the
Date
list:
•
Text Based — DD/MM/YYYY (16 Jan 2019)
•
Month First — MM/DD/YYYY (01/15/2019)
•
Day First — DD/MM/YYYY (15/01/2019)
•
Year First — YYYY/MM/DD (2019/01/16)
3.
Select the
12 hour am/pm
or
24 hour
(24-hour clock format, such as 13:22)
option to designate the time format.
4.
Select one of the following format options from the
Numbers
list:
•
123,456,789.00
•
123.456.789,00
To enable screen
features
1. Navigate to the
Local Settings
panel (
Settings
s
screen >
System Setup
tab).
2.
In the
Screen
area, select or de-select the Screensaver Enabled option. Click
the
Wait
field to use the numeric keypad to set the number of idle minutes
before the Analyzer initiates a screensaver.