Conductivity Value IEEE 32-bit float
Conductivity Value Low Word
3500
// 16-bit unsigned
Conductivity Value High Word
3501
// 16-bit unsigned
Conductivity Units
3502
// 16-bit unsigned
1 = nS/cm
2 = uS/cm
3 = mS/cm
Resistivity Value IEEE 32-bit float
Resistivity Value Low Word
3600
// 16-bit unsigned
Resistivity Value High Word
3601
// 16-bit unsigned
Resistivity Units
3602
// 16-bit unsigned
1 =
-cm
2 = k
-cm
3 = M
-cm
Temperature Value IEEE 32-bit float
Temperature Value Low Word
3700
// 16-bit unsigned
Temperature Value High Word
3701
// 16-bit unsigned
Temperature Units
3702
// 16-bit unsigned
1 =
C
2 =
F
3 =
K
Raw Conductivity Value IEEE 32-bit flat
Raw Conductivity Value Low Word
3800
// 16-bit unsigned
Raw Conductivity Value High Word
3801
// 16-bit unsigned
Raw ConductivityUnits
3802
// 16-bit unsigned
1 = nS/cm
2 = uS/cm
3 = mS/cm
Holding Registers (read/write)
not used
N
OTE
: For some PLCs and systems using Modbus RTU, you may need to offset the register by 1. For
example, 3300 would become 3301.
Setup Menu Options
Output Rate
The CheckPoint makes a TOC measurement every 15 seconds. The output rate can be changed to
select longer intervals between measurements, either by averaging the measurements that have
been taken in the last interval or by the last measurement for that interval. The output interval can be
set from 15 seconds to 8 hours.
Set the output rate as follows:
1. On the
Setup
screen, use the
Vertical Scroll
bar to highlight
Output Rate
, and then press
the
Enter
key.