SF-8200 Operation Manual
14
/
76
push the bar code scanning, and prompted to load the sample information in the software.
Warning: The sample has potential biochemical hazards, please wear rubber gloves
operation.
Warning: laser radiation, do not look directly at the light beam.
6) Reagent positions
Intermediate reagent area: 24 holes, can place 1 bottle of 25ml specifications reagent, 3 bottles of
15ml specifications reagent, 10 bottles of 5ml specifications reagent, 10 bottles of 2ml specifications
reagent, which contains two reagent stirring position, located in the middle reagent position
Start the reagent area: 18 holes, can place 3 bottles of 25ml specifications reagents, 15 bottles of
5ml specifications reagent, which contains three reagents stirring position, left in the start reagent posi-
tion three holes.
2 sets of independent cleaning tank, respectively, cleaning the sample needle and reagent needle.
With condensate collection and discharge function, to improve the reagent position in high temperature
and humidity environment stability.
7) test positions
Solidification test positions: 4
Optical method test positions: 4
Test zone heating function, for the sample during the test to maintain a constant temperature.
Warning: Beware of possible high temperatures on the surface of the test site. Use cau-
tion.
3.2 Specifications
1) temperature control
When the room temperature is 15
℃
~ 30
℃
, the temperature control in the following range:
Temperature and test position temperature control bit 37.0
℃
± 1.0
℃
;
Reagent temperature control does not exceed 16
℃
;
The preset temperature is reached within 30 minutes after the power is turned on.
2) test items and reporting units
Using the clotting method, chromogenic substrate method, immunoturbidimetric method for the
following coagulation tests and scalable items
Items
Clotting method
(Mechanical method)
Optical method
Reportable unit
chromogenic sub-
strate method
immunoturbidimetric
method
PT
※
s, INR, %, R
APTT
※
s