background image

Page 25

GENERAL TROUBLESHOOTING

Error

Possible Cause

Solution

Appliance will not start.

Fuse is defective.

Verify the fuse is not blown, damaged, or

improperly seated.

Mains cable (external) is defective.

Check the mains cable for visible damage or

pinches. Make sure the connections are

connected and tight.

Mains cable (internal) is defective.

Make sure the connections are connected

and tight.

Control panel is defective.

Check for visible damage to the control

panel.

Control board is defective.

Check.

Ribbon cable between the control board and

control panel electronics is defective.

Check the ribbon cable for twists, tears, or

pinches.

Ribbon cable between the control board and

control panel electronics is loose.

Check the ribbon cable for proper connector

seating.

Appliance is not being ventilated.

Mains cable (external) is defective.

Check the mains cable for visible damage or

pinches. Make sure the connections are

connected and tight.

Mains cable (internal) is defective.

Make sure the connections are connected

and tight.

Control panel is defective.

Check for visible damage to the control

panel.

Control board is defective.

Check.

Ribbon cable between the control board and

control panel electronics is defective.

Check the ribbon cable for twists, tears, or

pinches.

Ribbon cable between the control board and

control panel electronics is loose.

Check the ribbon cable for proper connector

seating.

Appliance is not being ventilated. Lid will not

open.

Power failure.

Check the electrical supply.

Relay on control board is defective.

Replace the control board.

Valve block cable/plug connector is

defective.

Check the cable for proper seating and

continuity.

Valve is defective.

Test the valve in the Service menu. If it does

not respond, replace the valve.

Sharp change in air pressure (range hood,

slammed door)

Unplug the power cable. Plug in the power

cable, and touch  .

Lid pressed in too much

Unplug the power cable. Plug in the power

cable, and touch  .

Stand-by (ON/OFF) is in order, but buttons

are unresponsive.

Automatic button calibration fault.

Wait 5 seconds.

Control panel is defective.

Replace the control panel.

Button does not function despite being

repeatedly selected.

When the appliance is turned on, verify there

are no damp cloths/objects/hands on the

operating panel. The buttons can short-

circuit each other.

1

Make sure the control panel is dry.

2

Remove any objects, and dry the control

panel.

3

Switch the appliance off and then back

on again.

Stand-by (ON/OFF) is in order—pump not

starting—display on control panel

electronics, as no significant detectable fall

in pressure after 5 seconds.

Vacuum pump is defective.

Test vacuum pump function.

Control board is defective.

Check and replace if necessary.

Control board relay is not switching.

Check and replace if necessary.

Control board to vacuum pump cabling is

defective/loose.

Check and replace if necessary.

Capacitor is defective (humming) or is not

connected.

Check and replace if necessary.

ICB Vacuum Seal Drawer

Troubleshooting Guide

TROUBLESHOOTING GUIDE 

service.subzero.com

826148 REV. A     01/2019

24

 

Summary of Contents for WOLF ICB

Page 1: ...n 2 Theory of Operation 3 Controls Operation 5 Component Access Removal 7 Troubleshooting Guide 25 Test Procedures 30 Technical Data 33 Wiring Diagrams 34 subzero com 800 222 7820 ICB Vacuum Seal Draw...

Page 2: ...e or in part without the express written consent of Wolf Appliance Inc Wolf Appliance Inc all rights reserved WOLF PRODUCT RATING PLATE INFORMATION On Vacuum Seal Drawers the product rating plate is o...

Page 3: ...he ventilation valve Air enters the vacuum chamber restoring it to ambient air pressure 5 The return to ambient air pressure causes the vacuum lifting cylinders to lower This opens the safety switch i...

Page 4: ...a reed switch that senses the magnet in the lid Lid and Gasket The glass lid is laminated and double glazed The lid and its gasket hermetically seal the vacuum chamber The view into the vacuum chamber...

Page 5: ...desired vacuum level 6 Touch multiple times to select the desired seal level 7 Press down on the lid and touch to initiate the vacuum process An audible signal indicates the process is complete 8 Ope...

Page 6: ...ecified vacuum level is not achieved within the required time about two minutes 10 times the appliance advises the user to do a pump dehumidification cycle The light illuminates From this point the us...

Page 7: ...s until it comes free 4 Remove the gasket from the unit GASKET INSTALLATION Install the Gasket Install the gasket into the channel in the lid When fitting the gasket make sure its seam is positioned b...

Page 8: ...lid 2 Pull the sealing bar assembly upward to disengage it from its electrical connections 3 Remove the sealing bar assembly from the vacuum chamber After removing the sealing bar assembly it is poss...

Page 9: ...ve the hinge cover from the unit HINGE COVER INSTALLATION Install the Hinge Cover 1 Place the hinge cover over the hinge 2 Push the hinge cover toward the rear of the unit until it clicks into place L...

Page 10: ...e HINGE SPRING REMOVAL To remove the hinge spring first remove 1 Both hinge covers 2 Lid Remove the Hinge Spring Remove the hinge spring from the lid HINGE SPRING INSTALLATION Install the Hinge Spring...

Page 11: ...ove it from the unit FRONT PANEL INSTALLATION Install the Front Panel 1 Attach adhesive strips to the front panel to prevent it from sliding 2 Push the front panel downward onto the front of the unit...

Page 12: ...CAUTION The drawer s vacuum pump contains oil that can spill if the drawer is tilted Whenever doing any service on the drawer make sure that it is level and stable If the vacuum pump oil spills it ca...

Page 13: ...Adjust the position of the drawer on the slides if necessary 4 Use a 5 5 mm hex driver to install the screws fixing the drawer to the slides on both the left and right CAUTION Avoid tipping the drawer...

Page 14: ...lides and pull it forward Then remove 1 Both side panels Remove the Vacuum Pump Cover 1 Use a T 15 Torx bit to remove the screws holding the vacuum pump cover to the rear of the unit 2 Remove the vacu...

Page 15: ...ins connection to control board Ground connection 2 Remove the cable ties on the bottom of the drawer 3 Feed the mains connection and ground connection through the hole in the side of the drawer so th...

Page 16: ...ng 3 Insert the housing into the cabinet 4 Install the screws between the housing s pilaster strips and the cabinet 5 Plug the power cord into the power socket Install the Vacuum Seal Drawer into the...

Page 17: ...vacuum pump contains oil that can spill if the drawer is tilted Whenever doing any service on the drawer make sure that it is level and stable If the vacuum pump oil spills it can contaminate the vac...

Page 18: ...ttom 2 Remove the capacitor from the unit 3 Mark the capacitor wire connectors 4 Disconnect the wire connectors between the capacitor and vacuum pump CAUTION High voltage may be present at the capacit...

Page 19: ...ontains oil that can spill if the drawer is tilted Whenever doing any service on the drawer make sure that it is level and stable If the vacuum pump oil spills it can contaminate the vacuum system or...

Page 20: ...s to the components behind the tray Make sure no cables or tubes are crushed or damaged Take particular care that the transformer cable the cable on the vacuum lifting cylinders for the safety switch...

Page 21: ...pump to the capacitor TRANSFORMER REMOVAL WARNING Turn off the power to the product before starting this procedure CAUTION The drawer s vacuum pump contains oil that can spill if the drawer is tilted...

Page 22: ...board 2 Remove the tube connection blue tube pressure sensor on control board to sealing bar valve 3 Loosen the four standoffs on the right hand side of the drawer 4 Remove the control board from the...

Page 23: ...sealing bar valve left to right black and blue vent valve left to right blue and brown VACUUM LIFTING CYLINDER REMOVAL WARNING Turn off the power to the product before starting this procedure CAUTION...

Page 24: ...enever doing any service on the drawer make sure that it is level and stable If the vacuum pump oil spills it can contaminate the vacuum system or the inside of the unit To remove the safety switch fi...

Page 25: ...Lid will not open Power failure Check the electrical supply Relay on control board is defective Replace the control board Valve block cable plug connector is defective Check the cable for proper seati...

Page 26: ...etry Appliance is vacuuming and sealing but too much air is being left in the bag container vacuum level 3 is unsatisfactory Through using the appliance moisture can get into the oil circuit and reduc...

Page 27: ...nt but not working Replace magnet and position correctly Magnet missing Attach magnet Magnet broken Replace Reed contact defective Replace reed contact Plug connection for reed contact is defective or...

Page 28: ...use new solenoid coils or replace the whole valve block Appliance aborts after 5 seconds and is making noises Gasket not tight Replace the gasket Leakage in tube system pump cannot be heard Replace th...

Page 29: ...r detergent to clean the appliance Appliance aborts after 5 seconds vacuum is not being created Gasket dirty or cracked Check Clean or replace if necessary Oil level of pump is at the minimum or 0 Oil...

Page 30: ...alve can be heard venting If the venting valve is in order the noise can be heard each time the button is touched TEST THE SEALING BAR VALVE In the Service menu touch the button The sealing bar valve...

Page 31: ...available to the appliance 2 Check the supply voltage for the control board 220 VAC between X1 pins 1 and 2 3 If the supply voltage for the control board is OK check the components supply voltages Vac...

Page 32: ...les energize the meter should read approximately 15 VAC If power is available and the sealing function does not work the sealing bar is not working Otherwise no power at the poles during sealing indic...

Page 33: ...reduction from ambient pressure Vacuum level 1 75 reduction from ambient pressure Vacuum level 2 85 reduction from ambient pressure Vacuum level 3 98 reduction from ambient pressure Sealing Levels Se...

Page 34: ...ICBVS24 WIRING DIAGRAM ICB Vacuum Seal Drawer Wiring Diagrams WIRING DIAGRAMS service subzero com 826148 REV A 01 2019 33...

Reviews: