Recirculating Cooler 

SRC14

Instructions for use

Version 2.0

Summary of Contents for SRC14

Page 1: ...Recirculating Cooler SRC14 Instructions for use Version 2 0 ...

Page 2: ...n provided may be impaired Electrical Installation THIS EQUIPMENT MUST BE EARTHED Before connection please ensure that the line supply corresponds to that shown on the rating plate This model requires a supply rated at 220 240V 50 Hz single phase Model Supply requirements Power SRC14 230V 50 Hz 1 1kW There is an IEC socket at the rear of the instrument for connection to the mains supply The unit i...

Page 3: ...sive force connecting the hoses use universal clips or similar to secure the tubing if necessary Always consider the temperature and pressure requirements before selecting your tubing material Use tubing of the minimum length required excessively long tubing will reduce the efficiency of the unit Filling procedure When choosing your cooling medium you should take care that there is no risk of corr...

Page 4: ... more than 5 seconds Pb will be displayed Using the up and down key to select the correct value as measured on the thermometer calibration can only be performed between 9 9ºC and 9 9ºC After setting the correct value press the set key for more than 5 seconds the display will return to the current measured temperature After operation Switch the circulation pump off then switch off the power switch ...

Page 5: ...rformance and safety features designed into the instrument If in any doubt please contact the Technical Service Department of Bibby Scientific Ltd or the point of sale To open fusebox insert a small screwdriver into the slot and push downwards Drainage hose Fuse box AC outlet Drainage hose Fuse box AC outlet Bibby Scientific Ltd Stone Staffordshire ST15 0SA United Kingdom Tel 44 0 1785 812121 Fax ...

Page 6: ...contact your local service representative Follow procedure to change fuse in the Maintenance and services section Contact local service representative Cooling power is poor Refrigeration unit does not work The ambient temperature exceeds 35ºC Exhaust is blocked Refrigeration air filter is blocked Contact local service representative Try to lower the ambient temperature Check necessary clearance ar...

Page 7: ...fications We cannot be sure that other equipment used in its vicinity will meet these standards and so we cannot guarantee that interference will not occur in practice Where there is a possibility that injury damage or loss might occur if equipment malfunctions due to radio frequency interference or for general advice before use contact the Technical Department of Bibby Scientific Ltd ...

Page 8: ...225 9243 Fax 609 589 2571 www bibby scientific com Bibby Scientific Italia Srl Via Alcide de Gasperi 56 20077 Riozzo di Cerro al Lambro Milano Italia Tel 39 0 2 98230679 Fax 39 0 2 98230211 e mail marketing bibby scientific it www bibby scientific it Bibby Scientific Asia Ltd Room 607 Yen Sheng Centre 64 Hoi Yuen Road Kwun Tong Kowloon Hong Kong Tel 852 3583 1581 Fax 852 3583 1580 e mail bibby bib...