PRO 360
Owner's Manual
Page 1: ...PRO 360 Owner s Manual ...
Page 2: ...latformthat allows the state of the art sensor and its microprocessor circuit to provide unsurpassed accuracy throughout the Protractor s 360 range The Pro 360 uses an innovative liquid filled angle sensor As the protractor is moved the liquid seeks a new position resulting in changes in the electrical properties of the sensor A microprocessor analyzes the changes and calculates the numbers seen o...
Page 3: ... display purposes In each the Pro 360 achieves an accuracy of 0 1 at level and plumb with a maximum error of 0 2 in between And the Pro 360 does not need to be returned to the manufacturer or dealer for recalibration By following the Superset instructions in this booklet you can test and recalibrate the Pro 360 in just a couple of minutes on site and without any special fixtures ON OFF button Liqu...
Page 4: ...ndisturbed for five minutes to reactivate push the ON OFF button The Pro 360 also indicates when the battery is low Change the battery when the display alternately flashes LO bAt with angle measure ments NOTE The Pro 360 does not display inaccurate angles due to a weak battery To install replace the battery Unscrew the battery compartment cover screw Remove the cover Install or replace the battery...
Page 5: ... noting the angle When the unit is first turned on the displayed angles will be in standard reference mode level true horizontal is displayed as 0 0 and plumb true vertical is displayed as 90 0 An alternate reference point can be set easily if desired see ALT ZERO An arrow on the left side of the display will indicate which way to move the Pro 360 to achieve level or plumb Between 0 and 44 9 the a...
Page 6: ...been in position and still for 5 seconds The readout will freeze and a flashing will appear in the right upper hand corner of the display An arrow on the left side of the LCD will also flash To release press the Hold button again Note The Pro 360 cannot be recalibrated Superset nor can a new reference point be established ALT ZERO when the HOLD feature is engaged The HOLD feature is disengaged aut...
Page 7: ... surface and wait 5 seconds Press the ALT ZERO button once Alt will appear on the display followed by 0 The Pro 360 will then display angles using the new reference A flashing will appear in the lower right hand corner of the display as long as the Pro 360 is in ALT ZERO mode Press ALT ZERO again to return to standard reference mode The Pro 360 also returns to standard reference when it is turned ...
Page 8: ...quire any special fixtures tools or expertise Accuracy Test Perform this simple test each day before using the Pro 360 Also perform it anytime the Pro 360 has been dropped or is being used in an environment that varies more than 5 C 9 F from the environment in which it was last calibrated If your Pro 360 fails this test you must recalibrate it using Superset before using the unit further to measur...
Page 9: ...splay is facing you but the lettering on the face of the unit is upside down Wait 10 seconds and note the angle on the display Finally rotate the unit end for end so the display is facing away from you the lettering should still be upside down Wait 10 seconds and note the angle on the display If any of the four measurements varies from any other by more than 0 1 you must recalibrate Superset your ...
Page 10: ...he same surfaces throughout the entire process Note Each time you reposition the Pro 360 during Superset wait a minimum of 10 seconds before pressing the HOLD button to advance to the next step Starting Superset Press and hold the HOLD and ALT ZERO buttons simultaneously Keep them depressed for approximately three seconds Release the buttons when the symbol SUP appears A 0 within flashing brackets...
Page 11: ... up Align with same edge or line wait 10 seconds Press HOLD button until 2 appears 3 Roll unit so it faces you the lettering should now be upside down Align with same edge or line wait 10 seconds Press HOLD button until 3 appears 4 Rotate unit so it faces away from you the lettering should still be upside down Align with same edge or line wait 10 seconds Press HOLD button until 4 appears You have ...
Page 12: ...ttom to top Align with same edge or line wait 10 seconds Press HOLD button until 6 appears 7 Rotate unit end for end so it faces you the lettering should now read top to bottom Align with same edge or line wait 10 seconds Press HOLD button until 7 appears 8 Roll the unit so it faces away from you the lettering should still read top to bottom Align with same edge or line wait 10 seconds Press HOLD ...
Page 13: ...strial use The following tips will ensure a long service life Use the end for end accuracy test daily to make sure the unit is in calibration If it is not recalibrate it immediately with Superset Clean the Pro 360 with mild liquid soap and a damp cloth Never immerse in water Do not use solvents directly on any of the Pro 360 plastic surfaces Store the Pro 360 away from extreme temperature Never be...
Page 14: ... Temperature Operating 5 C to 50 C 23 F to 122 F Storage 20 C to 65 C 4 F to 149 F Weight 325 g 11 5 oz Repeatability 0 1 Cross Axis Error Minimal Power 9 V Alkaline Battery Battery Life 500 hours typical Warranty 1 year limited 140 1 215 2 6 015 4 125 distance between two 6 32 tapped holes 1 22 6 04 14 ...
Page 15: ...ot perform as warranted herein the original purchaser s sole remedy will be the repair or replacement of the components as provided above In no event will the manufacturer be liable for damages lost revenue lost wages lost savings or any other incidental or consequential damages domestic or international rising fromthepurchaseanduseorinabilitytousethecomponentsevenifthemanufacturerhasbeenadvisedof...