Avoid locating equipment adjacent to the Surg System. If such a configuration is
necessary, verify normal operation of both the Surg System and the adjacent
equipment during use.
•
Always maintain a minimum separation distance of 6 inches [15 cm] from an in-use pacemaker
and the UHF RFID antenna of the Surg Reader while in use. Failure to comply may
cause interference with the pacemaker.
G
Gu
uiid
da
an
ncce
e a
an
nd
d M
Ma
an
nu
uffa
accttu
urre
err’’ss D
De
ecclla
arra
attiio
on
n -- E
Elle
eccttrro
om
ma
ag
gn
ne
ettiicc E
Em
miissssiio
on
nss
This equipment is intended for use in the electromagnetic environment specified below. The
customer or the user of this equipment should assure that it is used in such an environment.
E
Em
miissssiio
on
nss T
Te
esstt
C
Co
om
mp
plliia
an
ncce
e
E
Elle
eccttrro
om
ma
ag
gn
ne
ettiicc E
En
nvviirro
on
nm
me
en
ntt -- G
Gu
uiid
da
an
ncce
e
FCC
Part 15
W
WA
AR
RN
NIIN
NG
G -- Always maintain a minimum separation
distance of 9 inches [23 cm] from the head and torso of
a person and the UHF RFID antenna of the
Surg Reader while in use. Failure to comply
may cause FCC RF exposure limits to be exceeded.
The Surg System complies with FCC
regulations for conducted and radiated emissions under
FCC Part 15.
RF Emissions
CISPR 11
Group 2
W
WA
AR
RN
NIIN
NG
G -- The Surg System must emit
electromagnetic energy in order to perform its intended
function. Nearby electronic equipment may be affected.
RF Emissions
CISPR 11
Class A
W
WA
AR
RN
NIIN
NG
G -- This equipment is intended for use in a
professional healthcare facility environment. This
equipment may cause radio interference or may disrupt
the operation of other nearby equipment. It may be
necessary to take mitigation measures, such as
reorienting or relocating the Surg System or
shielding the location.
Harmonic
Emissions
IEC 61000-3-2
Class A
Voltage
Fluctuations/
Flicker
Emissions
IEC 61000-3-3
Complies
700001138880 Rev-AC