Hydraulic release pedals are not loose or damaged
Jack release valve is free of dust, debris, does not stick
Jack linkages are not out of adjustment or damaged
Jack adjustment valves and spring work
Jacks are not damaged
Head end and foot end jacks raise and lower at the same time
Litter raises and lowers from all locations
Litter components are in place and not damaged (fastener, holding pin, pin, bushing not backing out, loose, worn out,
or damaged)
Trendelenburg/Reverse Trendelenburg operates from all locations
Check skins for cracks
Hook and loop fastener is in place, intact, and secure
Fowler raises, lowers, and latches in place
Fowler subsystem (handle, wire, base weldment, cylinder, fasteners, etc. ) are not damaged
Hydraulic jacks are holding
No interference between wire and mechanical components on the Fowler backrest
No leaks at hydraulic connections
Lubricate where required
Push handles are not loose or damaged
Body restraints latch and are secure (optional)
IV pole is intact, is not damaged, and adjusts and latches in all positions (optional)
Oxygen bottle holder is intact and opens and closes (optional)
No rips or cracks in the mattress cover
Accessories and mounting hardware are in good condition
Product serial number:
Completed by:
Date:
R
Re
ettrra
acctta
ab
blle
e ffiifftth
h w
wh
he
ee
ell llu
ub
brriicca
attiio
on
n
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
MPG-3 grease
•
Bungee cords
P
Prro
occe
ed
du
urre
e::
1. Raise the product to the highest position.