80
4-0003045 Rev. G
After cleaning and/or disinfection, inspect all components for corrosion, damaged
surface, chips or contamination. Components found to be contaminated must be
cleaned and low-level disinfected again following the procedures described above.
Visually inspect the SPY-PHI imager, including its cables, for the following:
▪
Damage
▪
Sharp edges
▪
Loose or missing parts
▪
Rough surfaces
▪
Residue from cleaners and disinfectants (residues must be removed)
▪
Text and labels which are required for the safe and intended use must be
legible
WARNING:
Observe caution with a damaged and incomplete product (injury to the
patient, operator, or third parties is possible).
Perform a check before and after each use. Do not continue to use a product that
is damaged and/or incomplete or has loose parts. Contact Stryker Customer
Support to send in the damaged product with loose parts to be repaired. Do not
attempt to carry out any repairs on your own.
After cleaning and/or disinfection, store the product in sterile goods packaging as
follows:
▪
Protected from humidity and temperature fluctuations
▪
Protected from direct sunlight
▪
Protected from dust
Summary of Contents for SPY PHI HH9000
Page 1: ......
Page 6: ...iv 4 0003045 Rev G This page intentionally left blank...
Page 8: ...2 4 0003045 Rev G This page intentionally left blank...
Page 36: ...30 4 0003045 Rev G This page intentionally left blank...
Page 58: ...52 4 0003045 Rev G This page intentionally left blank...
Page 102: ...96 4 0003045 Rev G This page intentionally left blank...
Page 106: ...100 4 0003045 Rev G This page intentionally left blank...
Page 110: ...104 4 0003045 Rev G This page intentionally left blank...
Page 111: ......
Page 112: ......