Select Mean to set the brightness by weighting the adjustment more by the mean
brightness of the scene and less by the brightest points.
For example, use the Mean setting to set the brightness for viewing objects that are
farther from the front of the SPY-PHI imager. Foreground objects may appear too
bright to see detail.
Select Balanced to set the brightness by balancing the Peak and Mean settings.
For example, use the Balanced setting when viewing objects both near to
and
farther away from the front of the SPY-PHI imager. Balanced is the default setting
and is suitable for most scenarios.
Default Profile allows the user to return to the settings as originally installed on the
imaging system. The following are the default settings:
Table 9 Default Profile
Display Options
SPY
Overlay
SPY CSF (if available)
Image Quality
Sharpening: 4
Brightness: 192
Red Color Saturation: 768
Blue Color Saturation: 1024
Peak/Mean: Balanced
On-screen Info
Off
Summary of Contents for SPY PHI HH9000
Page 1: ......
Page 6: ...iv 4 0003045 Rev G This page intentionally left blank...
Page 8: ...2 4 0003045 Rev G This page intentionally left blank...
Page 36: ...30 4 0003045 Rev G This page intentionally left blank...
Page 58: ...52 4 0003045 Rev G This page intentionally left blank...
Page 102: ...96 4 0003045 Rev G This page intentionally left blank...
Page 106: ...100 4 0003045 Rev G This page intentionally left blank...
Page 110: ...104 4 0003045 Rev G This page intentionally left blank...
Page 111: ......
Page 112: ......