background image

SPR Plus® 

Overlay

 2790-100-000

Operations Manual

2021/05

2790-009-001 Rev AA.1

www.stryker.com

2021/05

2790-009-001 Rev AA.1

www.stryker.com

 WARNING

• 

Do not use this product with the bed siderails down. Siderails are to remind the patient of the 
edge of the unit, not as a patient restraining device.

• 

Do not reuse this product on another patient to avoid the risk of cross-contamination and 
infection.

• 

Always deflate the product before you perform CPR.

• 

Do not use this product as a transfer device.

• 

Always keep the bedding dry. Failure to keep the patient’s skin dry may lead to maceration.

• 

Do not continue to use the product if a bulge in the top surface is present. This may reduce 
the reliable product function.

• 

Always secure the product to the mattress with the corner straps.

• 

Always inspect the product for tears, cuts, holes, or any other damage, before use.

• 

Always check the patient’s skin at regular intervals for redness or breaks. If 

 

redness or breaks in the skin occur, consult the attending physician.

• 

Always consult the pump operations manual before connecting the system to this product.

 

CAUTION

• 

Improper usage of the product can cause injury to the patient or operator. Operate the 
product only as described in this manual.

• 

Do not modify the product or any components of the product. Modifying the product can 
cause unpredictable operation resulting in injury to patient or operator. Modifying the product 
also voids its warranty.

• 

Always use minimal layers of sheeting and incontinence pads. Too many layers between the 
patient’s skin and the product lowers the systems ability to aid in pressure reduction.

• 

Do not pull linens too tight over the product to avoid the hammock effect and reduce the 
effectiveness of the product.

• 

Always latch the siderails when the bed is in a raised position.

• 

Always be aware that the use of a mattress overlay may reduce the effectiveness and 
coverage of a siderail. 

Summary of safety precautions

Operation

Deflating for CPR

1. 

Unscrew the hose connector cap (Figure 3) from the product. 

        

2. 

Proceed per your hospital CPR procedures.

Note:

 The product will deflate in four to five seconds.

Stryker Medical

3800 E. Centre Avenue

Portage, MI 49002

USA

Figure 3: Unscrew hose 

connector cap

Reviews: