Cleaning
36
Hand wash all surfaces of the cot with warm water and mild detergent. Dry thoroughly. DO NOT STEAM
CLEAN, PRESSURE WASH OR ULTRASONICALLY CLEAN. Using these methods of cleaning is
not
rec-
ommended and may void this product’s warranty.
Thoroughly clean the cot once a month. Clean Velcro AFTER EACH USE. Saturate Velcro with disinfectant
and allow disinfectant to evaporate. (Appropriate disinfectant for nylon Velcro should be determined by the
hospital.)
Washing Procedure:
Foollow the cleaning solution manufacturer’s dilution recommendations exactly.
In general, when used in those concentrations recommended by the manufacturer, either phenolic type or
quaternary type disinfectants can be used. Iodophor type disinfectants are not recommended for use be-
cause staining may result.
Suggested cleaners for the 6092 cot surfaces:
Quaternary Cleaners (active ingerdient
−
ammonium chloride)
Phenolic Cleaners (active ingredient
−
o
−
phenylphenol)
Chlorinated Bleach Solution (5.25%
−
less than 1 part bleach to 100 parts water)
Avoid over
−
saturation and ensure the product does not stay wet longer than the chemical manufacturer’s
guidelines for proper disinfecting.
WARNING
SOME CLEANING PRODUCTS ARE CORROSIVE IN NATURE AND MAY CAUSE DAMAGE TO THE
PRODUCT IF USED IMPROPERLY. If the products described above are used to clean Stryker patient care
equipment, measures must be taken to insure the cots are wiped with clean water and thoroughly dried follow-
ing cleaning. Failure to properly rinse and dry the cots will leave a corrosive residue on the surface of the
cots, possibly causing premature corrosion of critical components.
NOTE
Failure to follow the above directions when using these types of cleaners may void this product’s warranty.
REMOVAL OF IODINE COMPOUNDS
Use a solution of 1/2 tablespoons Sodium Thiosulfate in a pint of warm water to clean the stained area.
Clean as soon as possible after staining occurs. If stains are not immediately removed, allow solution to
soak or stand on the surface. Rinse surfaces which have been exposed to the solution in clear water before
returning unit to service.
WARNING
Failure to properly clean or dispose of contaminated mattress or cot components will increase the risk of expo-
sure to bloodborne pathogens and may cause injury to the patient or the operator.
Summary of Contents for Rugged EZ-PRO R4 6092
Page 2: ......
Page 41: ...Notes 39 ...
Page 47: ...45 Maintenance Record Date Maintenance Operation Performed By ...
Page 49: ...Cot Assembly 47 6082 32 50 DETAIL 2 DETAIL 1 Assembly part number 6092 50 12 reference only ...
Page 50: ...Cot Assembly 48 Back of Fowler Foot Section Fowler Foot Section ...
Page 51: ...Cot Assembly 49 DETAIL B DETAIL C DETAIL A Detail A Detail C Detail B ...
Page 52: ...Cot Assembly 50 ...
Page 55: ...Base Assembly 53 Detail 3 Detail 1 Detail 2 Assembly part number 6092 1 11 reference only ...
Page 56: ...Base Assembly 54 DETAIL 1 DETAIL 2 DETAIL 3 ...
Page 57: ...Base Assembly 55 Apply stick wax ...
Page 58: ...Base Assembly 56 DETAIL 4 DETAIL 5 ...
Page 73: ...Litter Assembly 71 Assembly part number 6092 30 11 reference only ...
Page 74: ...Litter Assembly 72 DETAIL 1 DETAIL 2 ...
Page 75: ...Litter Assembly 73 DETAIL 3 DETAIL 4 DETAIL 5 ...
Page 76: ...Litter Assembly 74 DETAIL 6 ...
Page 115: ......