22
6500-509-001 REV B
www.stryker.com
Product Inspection
The condition of the cot is the responsibility of the owner. It is important that the cot is working properly before the
product is put into service. Have a qualified service technician use the following list and the operation instructions to
check the cot before the product is put into service.
The battery must be charged prior to checking the features and condition of the cot.
Item
Routine
Battery
Unpack batteries and charger
Charge battery according to SMRT Power System instructions (6500-009-101) or DeWALT
®
Battery System instructions
The power indicator LED, located at the foot end control enclosure of the cot, is solid green when the battery is fully
charged or has adequately charged battery power.
CAUTION
When charging a battery in an ambulance, locate the charger in an enclosed cabinet and out of patient reach during
transport.
After the battery is fully charged, inspect the cot for the following points:
Item
Routine
Page
Battery
Charge spare battery (if necessary) according to SMRT Pak or DeWALT
®
battery instructions
Install battery into foot end enclosure - power indicator LED operates
Ensure that the battery remains firmly secured
Release and remove battery from foot end enclosure
Reinstall battery into foot end enclosure
Hydraulics
Inspect motor mount - all fasteners secure
Check cylinder attachments at both ends - all fasteners secure
Inspect main cable - all connections secure
Inspect hoses and cylinder seal for leaks
Electronic
Controls
Check power indicator LED - charged
Extend cot to raised position
Verify ”jog” function operates smoothly
Lower to retracted position - cot secures in a mid-height position
(motor does not operate)
Determine and set ambulance vehicle load height
Check high speed retract
Extend cot to full height - no drift
Manual Back-up
Release
Verify the manual back-up release handle functions properly - adjust
accordingly
With the cot empty, check the raise/lower function
With the cot loaded with a minimum of 99 pounds (45 kg), check the raise/
lower function
With the cot loaded with a minimum of 99 pounds (45 kg), check the load/
unload function
Summary of Contents for Power-PRO XT 6500
Page 2: ......
Page 90: ...90 6500 509 001 REV B www stryker com Return To Table of Contents Cot Assembly ...
Page 91: ...www stryker com 6500 509 001 REV B 91 Return To Table of Contents Cot Assembly ...
Page 92: ...92 6500 509 001 REV B www stryker com Return To Table of Contents Cot Assembly ...
Page 93: ...www stryker com 6500 509 001 REV B 93 Return To Table of Contents Cot Assembly ...
Page 97: ...www stryker com 6500 509 001 REV B 97 Return To Table of Contents Base Assembly ...
Page 98: ...98 6500 509 001 REV B www stryker com Return To Table of Contents Base Assembly ...
Page 106: ...106 6500 509 001 REV B www stryker com Return To Table of Contents Notes ...
Page 118: ...118 6500 509 001 REV B www stryker com Return To Table of Contents Foot End Assembly ...
Page 124: ...124 6500 509 001 REV B www stryker com Return To Table of Contents Optional Gatch Assembly ...
Page 171: ......