Before Commencing Work
2
Before Commencing Work
Environmental conditions
Please familiarize yourself with the safety notes listed in the enclosed device
manual as well as local safety rules and regulations before commencing any
work.
• The ambient temperature must be between 10 °C and 40 °C.
• The relative humidity must be between 30% and 75%.
• Place the device on a level surface in a dry environment.
• Before operating the device, make sure the installation area is free of materials
subject to or the cause of electrostatic electricity, e.g., floor coverings.
2.1
Technical/Professional Requirements
All service tasks such as safety inspections (SI), maintenance, and repair work
must be performed by personnel trained and authorized by Stryker. The follow-
ing is required:
• Knowledge of all safety-relevant rules and regulations necessary to operate in-
sufflation devices.
• Knowledge of the current technical developments and the software status.
• Knowledge of electro technical and electronic devices and processes.
• Knowledge about the use of electrical test and measurement equipment.
• Good mechanical skills.
• Stryker service technician certificate
2.2
Tools and Test Equipment
Make sure the measuring instruments are calibrated and in perfect working or-
der before starting any maintenance tasks. The following tools and measuring in-
struments must be kept readily available:
• Multimeter: Voltage up to 1000 V AC/DC, electric current 10 A, with ohmmeter
and continuity tester
• ESD protective mat, antistatic wristband, cable
• Serial 9-pin 1:1 PC cable
• Manometer measurement range 0-100 mm Hg, measurement class 1.0
• Manometer measurement range 0-7 bar, measurement class 1.6
• Manometer measurement range 0-100 bar, measurement class 1.6
• Flow column range 0-50 l/min CO
2
, calibrated in slpm (standard liter per
minute) 24 °C/1013 mbar, max. pressure drop at 9l/min < 10 mbar
• Safety test measuring instruments/tools
• Screwdrivers: 2x Phillips (medium), 2x slotted (medium)
• Socket wrench SW5, SW5, 5, SW7, SW10
• Gas bottle (pressure >50 bar)
• Jaw wrench: SW5, 5, SW7, SW24
• Allen wrench: SW2, SW4
• Box wrench for M18x1
• Side nippers: Small
• Assembly wrench for tube connection
• Tweezers or pointed pliers
• Calibration adapter
• Device manual
• Service manual
• Windows 9x or later, Windows NT4 with Service Pack 2 or later
• IBM PC or compatible with CD drive
• Device software (CD)
• Internet access
• Medical Product Brochure (MPB)*
*localized