Replacing Components and Modules
51
EN
6.9
Uninstalling/Installing BAM Board
In case of malfunction (dark touch display, green LED no longer lights up, no LED
on board is illuminated), the BAM board must be replaced completely.
DANGER!
Disconnect device from power supply.
1. Open the device as described in chapter 3.6 Opening the Device .
2. Remove the ST4 plug
(3)
from the BAM board by pulling up.
3. Remove the ST1 plug
(2)
from the BAM board and the IFM board by pulling
parallel towards top.
4. Press to release the two board latches from the board and pull out the board
until it extends slightly towards top.
5. Remove the ST2 plug
(6)
and ST4 plug
(5)
from the BAM board by pulling ver-
tically.
6. Completely remove BAM board
(4)
from device.
7. Follow these steps in reverse for assembly.
Perform the following tests/updates after installing the new component:
• Software update via Global Services.net
• Basic function test
• Safety test (see chapter 4.1 Safety Test)
Fig. 6-18 Position of BAM board
(1)
2x board latches
(2)
Plug ST1/flat ribbon communica-
tion cable
(3)
ST3 plug/cable leading to power
supply unit
(4)
BAM board
(5)
ST4 plug/flat ribbon cable leading
to service interface
(6)
ST2 plug/cable leading to NET
board
(2)
(3)
(1)
(1)
(6)
(5)
(4)