Replacing Components and Modules
47
EN
6.6
Uninstalling/Installing Mains Socket with Fuse Holder
If the mains socket is damaged or malfunctions, the component must be re-
placed in its entirety. The mains socket module is supplied as a replacement mod-
ule including all screws, cables with plugs, line filter, and fuse holder.
DANGER!
Disconnect device from power supply.
1. Open the device as described in chapter 3.6 Opening the Device .
2. Cut the cable ties of the cables leading to the NET board.
3. Pull ST2 plug from NET board (see also Fig. 6-20 Detail A: Power switch/NET
board).
4. Remove ground wire connection cable
(5)
from bottom of case.
5. Remove the two fastening screws
(3)
from the outer rear panel and remove
the mains socket with the cables by pulling towards back.
6. Follow these steps in reverse for assembly.
The following tests must be performed after installing a new mains socket with
fuse holder:
• Basic function test
• Safety test (see chapter 4.1 Safety Test)
Fig. 6-12 Position of mains socket
(1)
Fuse holder
(2)
Mains socket
(3)
2x fastening screws
(1)
(2)
(3)
Fig. 6-13 Mains socket (rear side, inside)
(4)
Cable for mains socket
(5)
Grounding connection
(5)
(4)