34
EN
Service Menu
5.11
Resetting to Factory Defaults
This menu option must only be selected if the device is so severely mis-calibrated
that a correct calibration is no longer possible. Please note that selecting this
menu option will restore all factory settings. The device is then
not
calibrated and
cannot be used until recalibrated.
Prerequisites:
•Device in service mode
Confirm with
SAVE
Press
%
/
&
key until display depicts
Factory default
EXIT
Confirm with
SAVE
Factory default
Service
please wait
Factory default
Calibration success
Factory default
Factory default
Service
Display