78
Annual Inspection
EN
13
Annual Inspection
Measured values and tolerances
Each test conducted has to be documented with date and signature on the test
log.
The following measuring tools and resources were used by the manufacturer to
determine the listed measurements and tolerances:
An authorized service technician must check the device if the specified parame-
ters and tolerances are exceeded.
13.1
Safety Test
1. Perform a visual inspection. Make sure that
• the fuse corresponds with the specifications indicated by the manufactur-
er,
• labels and stickers on device are legible,
• the mechanical condition of the device allows for its safe use,
• the device is clean to ensure proper and safe functionality.
2. Measure leakage currents according to IEC 60601-1 / EN 60601-1.
3. Measure protective conductor resistance according to IEC 60601-1 / EN
60601-1. The protective conductor resistance is measured while device is
connected to the power supply. The max. value is 0.2
.
4. Measure the insulation resistance with 500-700 V DC. The min. value is
50 M
The electric strength with high voltage cannot be measured.
As an alternative, perform safety test according to IEC 62353 / EN 62353.
13.2
Basic Function Test (in High Flow Operating Mode)
1. Remove insufflation tube from device.
2. Use the On/Off switch to turn the device on. The device conducts a device
check. A short signal can be heard. Set to High Flow operating mode.
3. The factory default settings are 15 mm Hg for the nominal pressure and 3 l/
min for the nominal flow.
4. The following values are displayed:
Nominal pressure 15* [mm Hg]
Nominal gas flow 3* [l/min]
Actual pressure 0 [mm Hg]
Gas consumption 0.0 [l]
*These values correspond with the factory setting. If values in the configura-
tion menu were changed, these changed values are displayed.
5. Start insufflation: Press the
START
function field. The following values are dis-
played:
Actual pressure 0 [mm Hg]
Veress insufflation
is displayed. Streaming gas can be heard at the insuffla-
tion tube.
6. Select the max. nominal gas flow. The following values are displayed:
Nominal gas flow max. value [l/min]
Actual pressure 0 [mm Hg]
Insufflation: High Flow
is displayed. Streaming gas can be heard at the insuf-
Manometer
Range 0-100 mm Hg, error class 1.6
Syringe
60 ml
Silicone tube
8 mm x 2 m
T adapter
8-8-8 mm
Veress cannula
length 100 mm
opening diameter 1.4 mm,
inner cannula diameter 1.6 mm
Summary of Contents for PneumoSure
Page 6: ......
Page 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Page 104: ......
Page 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Page 108: ......
Page 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Page 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Page 210: ......
Page 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 308: ......
Page 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Page 408: ......
Page 409: ......