1. Unplug the pump from AC power.
2. Unplug the mattress tubing from the pump.
3. Place the pump with the front pump housing side down on a work surface.
4. Using a pick, remove and discard the bumper from the back pump housing (
5. Place and press the bumper in place.
F
Fo
oo
ott rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
T20 Torx driver
•
6 mm socket
•
Ratchet
P
Prro
occe
ed
du
urre
e::
1. Unplug the pump from AC power.
2. Unplug the mattress tubing from the pump.
3. Place the pump front down on a work surface.
4. Using a T20 Torx driver, remove the six screws that secure the back pump housing to the front pump housing. Save the
screws.
5. Using two hands, rotate the pump assembly to the right so the front is on your left.
6. Separate the front and back pump housing. Place the back pump housing down on the work surface.
7. Using a ratchet and a 6 mm socket, remove the six nuts that secure the foot to the front pump housing. Save the nuts.
8. Remove and discard the foot.
9. Reverse the steps to reinstall (
Pump housing rubber foot assembly 2874-007-022
10. Run and pass the diagnostic test.
11. Verify proper operation before you return the product to service.
F
Fiilltte
err rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
T20 Torx driver
P
Prro
occe
ed
du
urre
e::
1. Unplug the pump from AC power.
2. Unplug the mattress tubing from the pump.
3. Place the pump front down on a work surface.
4. Using a T20 Torx driver, remove and save the screw that secures the filter access door.
5. Remove and save the access door.
6. Remove and dispose of the filter.
7. Reverse the steps to reinstall (
8. Verify proper operation before you return the product to service.
2874-009-002 Rev A.1
9
EN