background image

Hydro

Set

Mixing Instructions

A Step-By-Step Guide

Indications for Use

HydroSet is a self-setting calcium phosphate cement indicated 
to fill bony voids or gaps in the skeletal system (i.e. extremities,
craniofacial, spine and pelvis). These defects may be surgically
created or osseous defects created from traumatic injury to the
bony. HydroSet is indicated only for bony voids or gaps that are not
intrinsic to the stability of the bony structure.

HydroSet cured in situ provides an open void/gap filler that can
augment provisional hardware (e.g. K. wires, plates, screws) 
to help support bone fragments during the surgical procedure. 
The cured cement acts only as a temporary support media and is

not intended to provide structural support during the healing process.

HydroSet is a self-setting calcium phosphate cement intended for
use in the repair of neurosurgical burr holes, contiguous craniotomy
cuts and other cranial defects as well in the augmentation or
restoration of bony contour in the craniofacial skeleton.

Note

For optimal performance store HydroSet between
18-22˚C for at least 1.5 hours prior to surgery.
Please see the pack insert for detailed information.

Step 1

Firmly click the Cement Delivery Syringe into
place in the inner blister tray so it is at an angled
position with the finger loops resting on the
shelves of the tray.

Step 2

Inject all the liquid from the glass syringe into
the mixing bowl.

Step 3

Vigorously mix the powder and solution in 
a circular motion for 45 seconds, ensuring 
that all of the solution has been distributed
throughout the powder.

Step 4

Transfer the HydroSet to the syringe by 
pouring it into the side of the funnel. Use 
spatula to assist.

Step 5

Remove pre-attached funnel by turning
counter-clockwise.

Step 6

Attach the supplied cannula by turning clockwise.
Please remove the plastic cannula cover.

Step 7

Screw the plunger rod into the plunger at the base
of the syringe, while keeping the cannula elevated.

Step 8

Position the syringe so the cannula is pointing
up and push the plunger rod to remove excess air.

STEP

TOTAL
TIME

1

3

4

5

6

7

8

2

Mix
45 sec

Load 
Syringe
1 min 
45 sec

Inject &
Sculpt
2 min

Harden
4 min

45 sec

2 min
30 sec

*4 min
30 sec

8 min
30 sec

* Sculpting beyond 4 min 

30 sec may inhibit the 
cement setting properties