Guidelines and manufacturer's statement - electromagnetic compatibility
41
EN
14.5
Guidelines and Manufacturer's Statement - Electromagnetic In-
terference Immunity
The device FLOSTEADY Arthroscopy Pump is intended for use in an electromag-
netic environment as described below. The user/operator of the FLOSTEADY Ar-
throscopy Pump should make sure the device is operated within such an
environment.
*Note: U
T
is the mains alternating voltage before applying the test levels.
Electromagnetic
interference
immunity tests
Test level
Compliance
Electromagnetic envi-
ronment guidelines
Discharge of static
electricity (ESD)
according to
IEC 61000-4-2
± 6 kV contact
discharge
± 8 kV air dis-
charge
In compli-
ance
Floors should be made
from wood or concrete
or covered with ceramic
tiles. If the floor cover-
ing consists of synthetic
material, the relative
humidity should be at
least 30%.
Electrical fast
transients / bursts
according to
IEC 61000-4-4
± 2 kV for AC
power lines
± 1 kV for input
and output lines
In compli-
ance
The quality of the sup-
ply voltage should be
the same as the voltage
of a typical business or
hospital environment.
Surges according
to IEC 61000-4-5
± 1 kV normal
mode voltage,
± 2 kV common
mode voltage
In compli-
ance
The quality of the sup-
ply voltage should be
the same as the voltage
of a typical business or
hospital environment.
Blackouts, brown-
outs, and fluctua-
tions of the power
supply according
to IEC 61000-4-11
< 5% U
T
* (> 95%
dip in the U
T
) for
½ cycle
In compli-
ance
The quality of the sup-
ply voltage should be
the same as the voltage
of a typical business or
hospital environment.
If the user/operator of
device requires the con-
tinuation of functional-
ity after power
interruptions/disrup-
tions, it is recom-
mended to supply the
device with power from
an uninterruptible
power supply.
40% U
T
(60% dip
in the U
T
) for 5
cycles.
70% U
T
(30% dip
in the U
T
) for 25
cycles.
< 5% U
T
(> 95%
dip in the U
T
)for
5 s
Supply frequency
magnetic field
(50/60 Hz) accord-
ing to IEC 61000-
4-8
3 A/m
In compli-
ance
Magnetic fields of the
mains power fre-
quency should comply
with the typical values
of business and hospi-
tal environments.
Summary of Contents for Flosteady Arthroscopy Pump 200
Page 6: ......
Page 8: ...EN 18 Warranty and Service 47 19 Appendix 49 19 1 Test Log 49 Index 50 ...
Page 31: ...User Menu 25 EN User Menu Overview ...
Page 55: ...Appendix 49 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Page 58: ...DE 18 Garantie und Kundendienst 50 19 Anhang 52 19 1 Testprotokoll 52 Index 53 ...
Page 83: ...Anwendermenü 27 DE Übersicht zum Anwendermenü ...
Page 108: ...52 Anhang DE 19 Anhang 19 1 Testprotokoll Datum Ergebnis Bemerkung Unterschrift ...
Page 137: ...Menu de l utilisateur 27 FR Plan d ensemble ...
Page 161: ...Annexe 51 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 163: ...Index alphabétique 53 FR U Utilisation clinique 7 V Valeurs de mesure et tolérances 37 ...
Page 164: ......
Page 166: ...NL 18 Garantie en Klantenservice 50 19 Bijlage 52 19 1 Testprotocol 52 Index 53 ...
Page 191: ...Gebruikersmenu 27 NL Overzicht bij het gebruikersmenu ...
Page 216: ...52 Bijlage NL 19 Bijlage 19 1 Testprotocol Datum Resultaat Opmerking Handtekening ...
Page 218: ...54 Index NL Voorschrift van de fabrikant 33 38 Voorzorgsmaatregelen 13 ...
Page 220: ...PL 18 Gwarancja i obsługa klientów 49 19 Aneks 51 19 1 Protokół testowy 51 Skorowidz 52 ...
Page 245: ...Menu użytkownika 27 PL Przegląd menu użytkownika ...
Page 269: ...Aneks 51 PL 19 Aneks 19 1 Protokół testowy Data Wynik Uwagi Podpis ...
Page 271: ......