4. Reverse steps to reinstall.
5. Verify proper operation before you return the product to service.
E
En
nd
drra
aiill cce
en
nttrra
all cco
ollu
um
mn
n a
asssse
em
mb
bllyy rre
ep
plla
acce
em
me
en
ntt
W
WA
AR
RN
NIIN
NG
G -- Always have only Stryker service technicians or service personnel trained by Stryker perform this procedure.
Failure to comply can result in serious damage to the bed and or or severe injury to users or patients.
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
#1 Phillips screwdriver
•
#2 Phillips screwdriver
•
Small screwdriver
•
1/2” socket
•
3/8” combination wrench
•
3/32” Allen wrench
•
Cordless drill
•
#2 Phillips bit
•
Adjustable supports (or equivalent)
P
Prro
occe
ed
du
urre
e::
N
No
otte
e -- Always specify if the product is equipped with the double safety lock when you order a central column assembly.
Proper warning labels are affixed to assemblies with out the double safety lock.
1. Remove the mattress.
2. Raise the rails to the highest position.
3. Apply the brakes.
4. Using a #1 Phillips screwdriver, remove the eight screws (A) that secure the bottom half of the lower plastic cover (B) to
the top half of the lower plastic cover (C) (Figure 27). Save the screws.
5. Insert a small screwdriver between the two halves of the lower plastic cover. Separate the two halves. Use caution to
avoid scratching the covers or damaging the upper cover snap pins.
1900-009-002 Rev AA.6
39
EN