S
Se
errvviicce
e
P
Po
ow
we
erriin
ng
g O
OF
FF
F tth
he
e p
pu
um
mp
p
Press the power button and hold for two seconds to power OFF the pump.
F
Fu
usse
e rre
ep
plla
acce
em
me
en
ntt
Fuse: F3.5AH250V
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
Slotted screwdriver
•
Digital multimeter
P
Prro
occe
ed
du
urre
e::
1. Unplug the power cord from the wall outlet.
2. Unplug the support surface tubing from the pump.
3. Place the pump face down on a work surface.
4. Using a slotted screwdriver, remove the fuse cover located below the power inlet. Save the fuse cover.
N
No
otte
e -- If you need to check a fuse, use a digital multimeter.
5. Remove and discard the fuse.
6. Reverse steps to reinstall.
7. Verify proper operation before you return the product to service.
B
Bu
um
mp
pe
err rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
Small slotted screwdriver
•
G
Go
oo
off O
Offff® Household Heavy Duty Remover
P
Prro
occe
ed
du
urre
e::
1. Unplug the power cord from the pump.
2. Unplug the gray AST and red CPR connector from the pump.
3. Using a small slotted screwdriver, scrape under each bumper to remove. Start from the corner of each bumper to
remove the top bumper (A), bottom left bumper (B), and bottom right bumper (C) from the back cover (D) (Figure 1).
Discard the bumpers.
EN
4