Complete this functional test procedure where indicated. Make sure that you pass all acceptance criteria.
P
Prro
occe
ed
du
urre
e:
1. Plug the power cord from the controller into a wall outlet.
2. Press the Power button.
Acceptance criteria:
Pump turns on.
3. Press the SPR Plus button.
Acceptance criteria:
Button is tactile and toggles to SPR Plus mode.
SPR Plus button lights.
Up/Down buttons light.
4. Press the Up/Down buttons. Continue to press the Up/Down buttons to test the pressure mmHg value settings.
Acceptance criteria:
The mmHg pressure value settings light up.
Button is tactile and increases the mmHg pressure value setting to the maximum value then sequences to the
lowest value.
5. Press the IsoFlex LAL button.
Acceptance criteria:
Button is tactile and toggles to IsoFlex LAL mode.
The Up/Down and SPR Plus lights turn off.