1. Unplug the power cord from the wall outlet.
2. Unplug the support surface tubing from the pump.
3. Place the pump face down on a work surface.
4. Using a T20 Torx driver, remove the six screws that secure the back pump housing to the front pump housing. Save the
screws.
5. Using two hands, rotate the pump assembly to the right so the front is to your left.
6. Separate the front and back pump housing. Place the back pump housing down on the work surface.
7. Grasp and remove the pressure hose at the PCBA.
8. Unplug the three cables from the PCBA.
9. Using a #2 Phillips screwdriver, remove the two screws that secure the support bracket that hold the PCB assembly and
screen to the front pump housing.
10. Using a #2 Phillips screwdriver, remove the three screws that secure the board to the PCBA assembly to the front pump
housing.
11. Unlock the keypad ribbon connector and remove the ribbon cable from the PCBA assembly.
12. Reverse the steps to reinstall (
13. Verify proper operation before you return the product to service.
K
Ke
eyyp
pa
ad
d m
me
em
mb
brra
an
ne
e rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
T20 Torx driver
•
#2 Phillips screwdriver
•
Pick
P
Prro
occe
ed
du
urre
e::
1. Unplug the power cord from the wall outlet.
2. Unplug the support surface tubing from the pump.
3. Place the pump face down on a work surface.
4. Using a T20 Torx driver, remove the six screws that secure the back pump housing to the front pump housing. Save the
screws.
5. Using two hands, rotate the pump assembly to the right so the front is to your left.
6. Separate the front and back pump housing. Place the back pump housing down on the work surface.
7. Grasp and remove the pressure hose at the PCB.
8. Unplug the three cables from the PCB.
9. Using a #2 Phillips screwdriver, remove the two screws that secure the support bracket. The support bracket holds the
PCB assembly and the screen to the front pump housing.
10. Using a #2 Phillips screwdriver, remove the three screws that secure the board to the PCB assembly to the front pump
housing.
11. Unlock the keypad ribbon connector and remove the ribbon cable from the PCB assembly.
2863-009-002 Rev AA.1
13
EN