3. Place the pump face down on a work surface.
4. Using a T20 Torx driver, remove the six screws that secure the back pump housing to the front pump housing. Save the
screws.
5. Using two hands, rotate the pump assembly to the right so the front pump housing is to your left.
6. Separate the front and back pump housing. Place the back pump housing down on the work surface.
7. Using a #2 Phillips screwdriver, remove the five screws that secure the back board to the back pump hosing. Save the
screws.
8. Remove and save the back board.
9. Using a #2 Phillips screwdriver, remove the four screws that secure the handles to the back pump housing. Save the
screws.
10. Remove and discard the handle.
11. Reverse the steps to reinstall (
Handle assembly - 2874-007-031
12. Verify proper operation before you return the product to service.
H
Ho
oo
okk rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
T20 Torx driver
•
Stubby #2 Phillips screwdriver
•
Pick
P
Prro
occe
ed
du
urre
e::
1. Unplug the power cord from the wall outlet.
2. Unplug the support surface tubing from the pump.
3. Place the pump face down on a work surface.
4. Using a pick, remove and discard the information label on the hook assembly.
5. Using a stubby #2 Phillips, remove the screw (under the label removed in step 4) that secure the hook assembly to the
back pump housing.
6. Using a T20 Torx driver, remove the six screws that secure the back pump housing to the front pump housing. Save the
screws.
7. Using two hands, rotate the pump assembly to the right so the front is to your left.
8. Separate the front and back pump housing. Place the back pump housing down on the work surface.
9. Unplug the power input cable and the power output cable from the power supply.
10. Using one hand to hold the fish paper and a stubby #2 Phillips screwdriver, remove the four Phillips screws and washers
that secure the power supply to the main frame. Save the screws and washers.
11. Remove the power supply to the side and fish paper.
12. Using a stubby #2 Phillips screwdriver, remove the four screws that secure the pump frame to the main frame and set
aside. Save the screws.
13. Using a stubby #2 Phillips screwdriver remove the six screws that secure the hook assembly to the back pump housing.
Save the screws.
14. Remove and discard the hook assembly.
15. Reverse the steps to reinstall (
16. Verify proper operation before you return the product to service.
EN
10
2863-009-002 Rev AA.1