Bed does not operate
1. Power cable does not plug to the
mains.
2. Power cable is damaged.
3. Nurse control unit or hand control
unit may be locked.
4. Control box does not operate
1. Ensure power cable and plug are
not damaged and then plug the
power cable to the mains.
2. Replace the power cable by the
authorized service personnel.
3. Nurse control Unit activation button
should be checked. (On-off).
Inspect the nurse control unit or
hand control unit and make sure
they are not locked.
4. Control box should be replaced,
call the technical service.
When the system operates, one of the
actuators does not operate and control
box gives a click sound
1. Actuator cable may be out of the
control box socket.
2. Actuator cable is damaged.
3. Actuator is damaged;
4. CPR wire remains stuck;
5. Control box is damaged.
1. Check socket connections of the
control box.
2. Actuator cable should be replaced,
call the service center.
3. Actuator should be replaced, call
the service center.
4. Inspect CPR wire and nut
adjustment.
5. Control box should be replaced,
call the service center.
When the system operates, one of the
actuators does not operate and control
box d
do
oe
ess n
no
ott give a click sound
1. Nurse control unit or hand control
unit may be partly locked.
2. Control unit is damaged,
3. Control box is damaged.
1. Inspect nurse control unit or hand
control and make sure they are not
locked.
2. Unplug side rail control unit from
the socket and check for function
with the nurse control unit. If
function is restored, their may be a
problem with one of the control
units. Call technical service.
3. Control box should be replaced,
call the service center.
Battery is possibly discharged and
does not give click sound
1. Battery is completely discharged
2. Battery is damaged or expired.
1. Recharge the battery.
2. Battery should be replaced, call the
service center.
System operates but actuators move
slowly
Power is possibly disconnected and
the bed operates by battery.
1. Power cable might have come out
of the control panel; check it.
2. Check the connections of the
power cable.
3. Check the socket.
4. Recharge the battery, call the
service center.
SK-7600 REV 06
15
EN
Summary of Contents for 7600-000-050
Page 2: ......
Page 16: ...S Se er riia all n nu um mb be er r llo oc ca at tiio on n A EN 12 SK 7600 REV 06 ...
Page 21: ...A B C D E F SK 7600 REV 06 17 EN ...
Page 68: ...H HM M 1 17 7 3 31 14 4 A A A HM 17 314 1 EN 64 SK 7600 REV 06 ...
Page 69: ...H HM M 1 17 7 1 16 6 A A HM 17 16 1 SK 7600 REV 06 65 EN ...
Page 70: ...H HM M 1 17 7 1 15 56 6 A A A HM 17 156 1 EN 66 SK 7600 REV 06 ...
Page 71: ...H HM M 1 17 7 3 31 10 0 H HM M 1 17 7 3 31 11 1 A HM 17 310 HM 17 311 1 SK 7600 REV 06 67 EN ...
Page 72: ...H HM M 1 17 7 3 31 13 3 A A HM 17 313 1 EN 68 SK 7600 REV 06 ...
Page 73: ...H HM M 1 17 7 3 31 12 2 A A HM 17 312 1 SK 7600 REV 06 69 EN ...
Page 74: ...H HM M 1 17 7 3 30 04 4 A B A HM 17 304 1 B HM 17 304 1 EN 70 SK 7600 REV 06 ...
Page 75: ...H HM M 1 17 7 3 30 03 3 A A A HM 17 303 1 SK 7600 REV 06 71 EN ...
Page 76: ...H HM M 1 17 7 3 30 05 5 A A HM 17 305 1 EN 72 SK 7600 REV 06 ...
Page 77: ......
Page 78: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA SK 7600 REV 06 WCR AA 11 2021 08 ...