Summary of Contents for 6297-000-000

Page 1: ...2019 09 6297 001 700 Rev AB www stryker com Instructions For Use ENGLISH EN Electric System 6 Sternum Saw 6297 000 000 ...

Page 2: ...duction This Instructions For Use manual is the most comprehensive source of information for the safe and effective use of your product This manual may be used by in service trainers physicians nurses surgical technologists and biomedical equipment technicians Keep and consult this reference manual during the life of the product The following conventions are used in this manual A WARNING highlight...

Page 3: ...s oxide ALWAYS place the handpiece in the safe mode position while the handpiece is idle before installing or removing any accessory and when passing the handpiece to another person ALWAYS operate the equipment within the specified environmental condition values See the Specifications section Accessories WARNING Use only Stryker approved system components and accessories unless otherwise specified...

Page 4: ... Critical product information may be lost if the cutting accessory is repackaged Upon initial receipt and before use visually inspect the package for damage to confirm the integrity of the sterile barrier Do not use the product if damage is apparent the sterile barrier is compromised or the package is unintentionally opened NOTE Sterile cutting accessories are sterilized by irradiation DESCRIPTION...

Page 5: ...e of two positions The teeth must always be directed away from the sternum blade guard Sternum Blade Guard Protects soft tissue from coming into contact with the blade Guard Actuating Knob To release the sternum blade guard push on knob with a circular motion Trigger The trigger is pressure sensitive for variable speed operation Applied Part The distal end of the handpiece and the sternum blade gu...

Page 6: ...ent inadvertent operation of the handpiece the handpiece cannot be operated Definitions The symbols located on the equipment and or labeling are defined in this section or in the Symbol Definition Chart See the Symbol Definition Chart supplied with the equipment SYMBOL DEFINITION Run Mode Safe Mode General warning sign Per European Union Directive 2012 19 EU product must be collected separately Do...

Page 7: ...ly tug the cutting accessory to ensure it is secure To Install Sternum Blade Guard WARNING DO NOT allow the blade to retract or extend from the end of the sternum blade guard NOTE The sternum blade guard may be installed in the orientation shown in figure 2 or rotated 180 degrees ALWAYS orient the sternum guard such that the blade teeth point away from the guard as illustrated Knob on Collar Figur...

Page 8: ... accessory pieces at a high velocity ALWAYS follow the recommended duty cycle to prevent the equipment from overheating See the Specifications section CAUTIONS When operating the handpiece let the blade do the cutting Applying too much pressure will bend the blade and reduce the cutting quality DO NOT stall the handpiece Failure to comply may damage the electric motor If the handpiece jams release...

Page 9: ...collar Remove the sternum blade guard from the handpiece Release the guard actuating knob to allow the spring loaded collar to snap back to its original position To Remove Cutting Accessory 1 Slide the run safe control to the safe mode position 2 Rotate the spring loaded blade collar clockwise to realign the slots and remove the blade Care Instructions For processing instructions and disposal recy...

Page 10: ...andpiece for repair Blade will not fit into the blade collar Debris is inside the end of the blade collar Clean the handpiece with a small brush Blade collar is damaged Return the handpiece for repair Blade collar does not spring back to retain the blade Debris is inside of blade collar Clean the handpiece with small brush Actuate the blade collar several times to obtain smooth operation Sternum b...

Page 11: ...without sternum blade guard Speed 14 000 cycles per minute Excursion 0 152 inch 3 9 mm Mode of Operation Duty Cycle Rest Between Cycles Non continuous Operation 1 minute on 4 minutes off 5 times 3 hours Power Supply Externally supplied CORE console REF 5400 050 000 CORE 2 console REF 5400 052 000 Ingress Protection IPX0 Ordinary Equipment Maximum Temperature of Applied Part Less than 124 F 51 C Ma...

Page 12: ... 700 Rev AB www stryker com Stryker Instruments 1941 Stryker Way Portage Michigan USA 49002 1 269 323 7700 1 800 253 3210 ES DE FR IT NL 6297 001 710 JA ZH KO 6297 001 720 SV DA FI PT NO 6297 001 730 PL EL 6297 001 750 ...