24
Operation
Note:
Before operating the device, ensure all system components have been
set up according to the instructions in the Setup section.
Powering the Console On/Off
Before using the camera in a surgical procedure, test all system
components to ensure proper function. Ensure that a video
image appears on all display monitors before beginning any
procedure.
1. Power on the display monitor.
2. Press the power switch on the console to power the console on or off.
Note:
A color bar pattern will appear on the display monitor if the camera
head is not connected to the camera console. If the color bar appears, refer to
the Setting Up the Camera Head section to connect the camera head.
Performing the White Balance Test
Before each surgical procedure, perform the White Balance test to adjust the
camera’s perception of white so it can display other colors correctly.
When a camera head is connected to the console and the console is powered
on, the display monitor will automatically prompt the user to perform the
White Balance test.
1
Using the camera head buttons, follow the instructions
on the display monitor to perform the test.
1
English must be selected as the language in the Advanced Settings.
The White Balance test can also be performed after the camera head is already
connected by pressing the WB button on the Home screen of the console (or
a camera head button if it has been programmed for White Balance).