ITEM:
TroubleShooting Guide
Record/Picture Switch
Increase/Decrease Brightness
Stryker 1188 HD Camera Head and Coupler
Camera Coupler
Focus Knob
1188 HD Camera Head
Unscrew soaking cap of the camera cord and locate the
Blue Arrow
on camera
cord. Plug the camera cord into the
Camera Head
with the blue arrow facing
Up
.
Make sure the camera is focused, and wipe the scope end and coupler window clean. Ensure
the coupler is clean and attached securely to the camera as moisture may be present from
sterilization. Re-White Balance the camera if needed.
Attach scope and light cord to camera. Ensure coupler has latch at
12 O’Clock
position, rotate clockwise if needed.
Rotate
Focus Knob
to focus and set depth after insertion. Press
Up
and
Down
arrows to increase or decrease
Brightness.
Zoom/White Balance Adjust
Activate light source, press and
Hold W Button
to white balance while pointed at gauze. Hold until
White Balance Complete
shows on monitor upper corner.
A Quick W Press
will cycle through the
Zoom
.
Blurry Image
Increase the light source power on the console. Press the
Up
arrow on the camera head to
increase the shutter as needed. Check light cables for broken fibers and replace if several black
spots are present when holding cable end up to a light.
Low Light
Ensure the
Scope end
and the
Coupler
attachment are clean of debris. Re-
White Balance
the
camera and try again. Make sure the camera is focused and
Decrease
the light source
Bright-
ness
and
Zoom
if needed. Make sure the light cables are in good condition.
Grainy Image
Decrease the light source power on the console. Press the
Down
arrow on the camera head to
decrease the shutter as needed.
Excess Glare
Common Issues and Troubleshooting