ITEM:

TroubleShooting  Guide

Record/Picture Switch

Increase/Decrease Brightness 

Stryker 1188 HD Camera Head and Coupler

Camera Coupler

Focus Knob

1188 HD Camera Head

Unscrew soaking cap of the camera cord and locate the 

Blue Arrow

 on camera 

cord.  Plug the camera cord into the 

Camera Head

 with the blue arrow facing 

Up

.

Make sure the camera is focused, and wipe the scope end and coupler window clean.  Ensure 
the coupler is clean and attached securely to the camera as moisture may be present from 
sterilization.  Re-White Balance the camera if needed.

Attach scope and light cord to camera.  Ensure coupler has latch at 

12 O’Clock 

position, rotate clockwise if needed.

Rotate 

Focus Knob

 to focus and set depth after insertion.  Press 

Up

 and 

Down

 

arrows to increase or decrease 

Brightness.

Zoom/White Balance Adjust

Activate light source, press and 

Hold W Button 

to white balance while pointed at gauze.  Hold until 

White Balance Complete

 shows on monitor upper corner.  

A Quick W Press

 will cycle through the 

Zoom

.

Blurry Image

Increase the light source power on the console.  Press the 

Up

 arrow on the camera head to 

increase the shutter as needed.  Check light cables for broken fibers and replace if several black 
spots are present when holding cable end up to a light.

Low Light

Ensure the 

Scope end

 and the 

Coupler

 attachment are clean of debris.  Re-

White Balance

 the 

camera and try again.  Make sure the camera is focused and 

Decrease

 the light source 

Bright-

ness

 and 

Zoom

 if needed.  Make sure the light cables are in good condition.

Grainy Image

Decrease the light source power on the console.  Press the 

Down

 arrow on the camera head to 

decrease the shutter as needed.

Excess Glare

Common Issues and Troubleshooting