6.
Move the adjustment ring up until the pin
fits into the V-shaped groove on the cabinet
of the control panel.
7.
Tighten the adjustment ring to fix it in this
position.
8.
Press the
Turn/push knob
knob to clear
the error message.
9.
Press the
Lower/Raise
button to raise the
specimen mover head.
5.7.3
Adjusting the horizontal position of the flexible specimen holder
With a flexible specimen holder:
1.
Press the
Lower/Raise
button to lower the
specimen mover head.
2.
Loosen the 2 fastening screws holding the
column of the control panel.
3.
Place the flexible specimen holder in a
position that does not allow the specimen
to run more than 1 mm over the edge of the
preparation disc.
5.8
LaboDoser-100 with LaboForce-100
If you are using LaboDoser-100 with LaboForce-100, see the specific machine manual.
5.9
LaboDoser-10 with LaboForce-100
If you are using LaboDoser-10 with LaboForce-100, a table
stand is required.
5.10
Noise and vibration levels
For information about noise and vibration levels, see the specific machine manual.